Agilent Technologies E1364A service manual Test 2-1 Closed-Channel Resistance Test

Page 23

Test 2-1: Closed-Channel Resistance Test

The purpose of this test is to verify that all relay contacts meet the closed-channel resistance specification for the switch module. If the closed-channel resistance of any contact is greater than 3.5 , the relay should be replaced.

Normally Open

Contacts 1. Hardware Connections

Figure 2-2. Closed-Channel Resistance (NO channels)

2.Equipment Setup

Set DMM to: 4-wire ohms, autorange.

Send *RST to the switch to open all NO contacts.

3.Closed-Channel Reading (channel 00)

Send CLOS (@100) to the switch to close channel 00.

Trigger the DMM and record the reading.

4.Open-Channel Reading (channel 00)

Send OPEN (@100) to the switch to open channel 00.

Trigger the DMM and verify that an open circuit is indicated (>108 ).

5.Closed- and Open-Channel Readings (channels 01-15)

Repeat steps 3 and 4 for channels 01 through 15. Use

CLOS (@ccnn) and OPEN (@ccnn), where

cc= switch card number (01-99, leading zero not necessary) nn = channel number (00-15)

Chapter 2

Verification Tests 23

Image 23
Contents Serial Numbers Agilent E1364A Channel Form C Switch ModulePage Contents Appendix B. Backdating Information Exclusive Remedies WarrantyLimitation Of Warranty CertificationDocumentation History Safety SymbolsModel Number Manufacturer’s NameManufacturer’s Address Declares, that the product Product NamePage Page Page Introduction Agilent E1364A Form C Switch ModuleGeneral Information Chapter Safety ConsiderationsChapter General Information Initial Incoming Inspection Guidelines Inspection/ShippingPackaging/Shipping Guidelines Environment Switch DescriptionTemperature Humidity Requirements Recommended Model Use Recommended Test EquipmentGeneral Information Chapter Verification Tests Operation Verification Functional VerificationPerformance Verification 1a. Agilent E1364A Test Fixture Schematic1b. Agilent E1364A Test Fixture Test 2-1 Closed-Channel Resistance Test Normally Closed Resistance Test Next 530 540 END Test 2-2 DC Isolation Test Normally Closed to Chassis 1. Hardware Connections DC Isolation Test NC Channels to Chassis Example DC Isolation Test Verification Tests Chapter Test Accuracy Ratio TAR Performance Test RecordPerformance Test Record for the E1364A Switch Module Page 1 Performance Test Record for the E1364A Switch Module Page 2 Maximum Uncert Test Measured Description MinimumTest Measured Description Minimum Reading Maximum Uncert Performance Test Record for the E1364A Switch Module Page 4Verification Tests Chapter Number Replaceable Parts ListsPart Qty Part Description Replaceable Parts Chapter E1364A Terminal Block Replaceable Parts Replaceable Parts Chapter Service Identifying the Problem Troubleshooting TechniquesE1364A Tests/Checks Test/Check Reference Designator Testing the AssemblyESD Precautions Repair and Maintenance GuidelinesFunctional Verification Test Verification Tests C ProgramsExample Closed Channel Resistance Test Performance Verification TestsClose instrument session Example DC Isolation Test Iclose dm Example C Programs Appendix a Reference Part Qty Part Description Designator Number Replaceable Parts ListQty Part Description Part NumberAgilent E1364A Reference Designators Figure B-1. E1364A Component Assembly Parts Figure B-2. E1364A Terminal Module Replaceable Parts Page Page Copyright Agilent Technologies, Inc