Agilent Technologies E1364A service manual Normally Closed to

Page 28

Normally Closed to

Common 1. Hardware Connections

Figure 2-5. DC Isolation Test (NC Channels to Common)

2.Equipment Setup

Set DMM to: 2-wire ohms, 1 Grange.

Send *RST to the switch.

Send the CLOS (@100:115) command to the switch to open all NC contacts.

3.DC Isolation Reading

Trigger the DMM and record the reading.

28 Verification Tests

Chapter 2

Image 28 Contents
Agilent E1364A Channel Form C Switch Module Serial NumbersPage Contents Appendix B. Backdating Information Warranty Limitation Of WarrantyCertification Exclusive RemediesSafety Symbols Documentation HistoryManufacturer’s Name Manufacturer’s AddressDeclares, that the product Product Name Model NumberPage Page Page Agilent E1364A Form C Switch Module IntroductionSafety Considerations General Information ChapterChapter General Information Inspection/Shipping Initial Incoming Inspection GuidelinesPackaging/Shipping Guidelines Temperature Humidity Switch DescriptionEnvironment Recommended Test Equipment Requirements Recommended Model UseGeneral Information Chapter Verification Tests Functional Verification Operation Verification1a. Agilent E1364A Test Fixture Schematic Performance Verification1b. Agilent E1364A Test Fixture Test 2-1 Closed-Channel Resistance Test Normally Closed Resistance Test Next 530 540 END Test 2-2 DC Isolation Test Normally Closed to Chassis 1. Hardware Connections DC Isolation Test NC Channels to Chassis Example DC Isolation Test Verification Tests Chapter Performance Test Record Test Accuracy Ratio TARPerformance Test Record for the E1364A Switch Module Page 1 Performance Test Record for the E1364A Switch Module Page 2 Test Measured Description Minimum Maximum UncertPerformance Test Record for the E1364A Switch Module Page 4 Test Measured Description Minimum Reading Maximum UncertVerification Tests Chapter Part Qty Part Description Replaceable Parts ListsNumber Replaceable Parts Chapter E1364A Terminal Block Replaceable Parts Replaceable Parts Chapter Service Troubleshooting Techniques Identifying the ProblemTesting the Assembly E1364A Tests/Checks Test/Check Reference DesignatorRepair and Maintenance Guidelines ESD PrecautionsVerification Tests C Programs Functional Verification TestPerformance Verification Tests Example Closed Channel Resistance TestClose instrument session Example DC Isolation Test Iclose dm Example C Programs Appendix a Replaceable Parts List Reference Part Qty Part Description Designator NumberPart Number Qty Part DescriptionAgilent E1364A Reference Designators Figure B-1. E1364A Component Assembly Parts Figure B-2. E1364A Terminal Module Replaceable Parts Page Page Copyright Agilent Technologies, Inc