Agilent Technologies E1364A service manual Functional Verification, Operation Verification

Page 20

Functional Verification

The Functional Verification Test for the E1364A switch consists of sending the *IDN? command and checking the response. This test can be used at any time to verify that the switch is connected properly and is responding to basic commands.

Procedure 1. Verify that the switch is installed in the mainframe and that the mainframe has passed its power-on test.

2.Send the *IDN? command to the switch module (see example following).

3.The switch module should return the following string (revision number may vary):

HEWLETT-PACKARD,SWITCHBOX,0,A.06.00

NOTE If the primary address setting, secondary address setting, or interface select code is set incorrectly, the switch will not respond. Verify proper address selection before troubleshooting.

Example An example follows which uses an HP 9000 Series 300 computer with HP BASIC and a switch address of 70915.

10

DIM A$[80]

 

20

OUTPUT 70915;"*IDN?"

!Send the ID command

30

ENTER 70915;A$

!Get response

40PRINT A$

50END

Operation Verification

The procedures in this section are used to provide a high confidence that the switch module is meeting published specifications. The Operation Verification tests are a subset of the Performance Verification tests and are suitable for checkout after performing repairs.

Operation Verification is performed by completing the Closed-Channel Resistance Test as described in the Performance Verification test procedures. This test is usually sufficient to verify that the switch module is meeting its specifications.

20 Verification Tests

Chapter 2

Image 20
Contents Agilent E1364A Channel Form C Switch Module Serial NumbersPage Contents Appendix B. Backdating Information Warranty Limitation Of WarrantyCertification Exclusive RemediesSafety Symbols Documentation HistoryManufacturer’s Name Manufacturer’s AddressDeclares, that the product Product Name Model NumberPage Page Page Agilent E1364A Form C Switch Module IntroductionSafety Considerations General Information ChapterChapter General Information Inspection/Shipping Initial Incoming Inspection GuidelinesPackaging/Shipping Guidelines Environment Switch DescriptionTemperature Humidity Recommended Test Equipment Requirements Recommended Model UseGeneral Information Chapter Verification Tests Functional Verification Operation Verification1a. Agilent E1364A Test Fixture Schematic Performance Verification1b. Agilent E1364A Test Fixture Test 2-1 Closed-Channel Resistance Test Normally Closed Resistance Test Next 530 540 END Test 2-2 DC Isolation Test Normally Closed to Chassis 1. Hardware Connections DC Isolation Test NC Channels to Chassis Example DC Isolation Test Verification Tests Chapter Performance Test Record Test Accuracy Ratio TARPerformance Test Record for the E1364A Switch Module Page 1 Performance Test Record for the E1364A Switch Module Page 2 Test Measured Description Minimum Maximum UncertPerformance Test Record for the E1364A Switch Module Page 4 Test Measured Description Minimum Reading Maximum UncertVerification Tests Chapter Number Replaceable Parts ListsPart Qty Part Description Replaceable Parts Chapter E1364A Terminal Block Replaceable Parts Replaceable Parts Chapter Service Troubleshooting Techniques Identifying the ProblemTesting the Assembly E1364A Tests/Checks Test/Check Reference DesignatorRepair and Maintenance Guidelines ESD PrecautionsVerification Tests C Programs Functional Verification TestPerformance Verification Tests Example Closed Channel Resistance TestClose instrument session Example DC Isolation Test Iclose dm Example C Programs Appendix a Replaceable Parts List Reference Part Qty Part Description Designator NumberPart Number Qty Part DescriptionAgilent E1364A Reference Designators Figure B-1. E1364A Component Assembly Parts Figure B-2. E1364A Terminal Module Replaceable Parts Page Page Copyright Agilent Technologies, Inc