Calibration
To enter the "calibration" mode, press any key at the same time as switching on the instrument.
Press the key for scale A or the key for scale B.
A number between 5,000 and 50,000 is displayed.
If this is not the case, the scale must be
If the display is stable, press the key to set the scale to zero.
Then place the calibration weight (3/6 kg = 500 g, 30 kg = 1000 g) on the platform.
(display 200000 ... 800000 at max. weight).
Enter the value of the calibration weight (3/6 kg = 500 ,
30 kg = 1) and press the key to calibrate the scale.
Switch off the scale, remove the load and switch the scale on again.
The scale is ready for use.
Error messages
Error | Remedy |
Max. weighing range | Remove weighing product |
exceeded. | from scale! |
|
|
Weighing platform not cor- |
|
rectly mounted. Scale outsi- | Mount platform correctly! |
de |
|
Set to zero not possible.
Make sure that the scale is empty!
Battery almost empty. | Connect the power plug! |