Teledyne 2101 installation and operation guide Volume 1 status code Bit integer Read

Page 50

2101 Field Wizard

Section 4 Modbus Protocol

Table 4-2 Modbus ASCII Address 2-(N+1) Register Definitions (Continued)

Register Number(s)

Name

Data Type

Units

Read/Write

(1)A write to the Identify module register will cause the module to perform the identify operation which may be a steady LED for a few seconds or a beep in the Field Wizard.

(2)Setting the Take Reading flag to 1 will cause the module to update the registers with current data readings. It will be set to zero when the readings have all been updated. This may be used to initiate readings and poll for when they are ready to be read. It may take up to 50 seconds to update all the readings, depending upon the flow con- ditions. Setting the Take Reading flag to 2 causes an automatic, 15 second update of readings when a Modbus master is polling the 2100.

(3)The Update Interval specifies an interval in seconds that the registers are automatically updated. It defaults to zero, which indicates that no automatic updating will occur.

(4)The Active Flag (1-4) bit fields specify what fields/registers are active in the list. This provides support for a maxi- mum of 64 fields. For example, if bit 0 of register 27 is set, the Level (registers 40,41) is active. If bit 1 of register 27 is set, then the Velocity (registers 55,56) is active. If bit 0 of register 28 is set, the Analog channel 7 (registers 265,266) is active.

(5)A non-zero status code indicates a measurement problem.

(6)Time is represented in a series of registers: Order is from lowest address to highest - Seconds (0-59), Minutes (0-59), Hours (0-23), Days (1-31), Month (1-12) and Year (1977-2099).

117

Volume 1 status code

16-bit integer

 

Read

 

 

 

 

 

118-123

Volume 1 time record

Time

 

Read

 

 

 

 

 

130,131

Voltage

4-byte float

Volts

Read

 

 

 

 

 

132

Voltage status code

16-bit integer

 

Read

 

 

 

 

 

133-138

Voltage time record

Time

 

Read

 

 

 

 

 

145,146

Temperature

4-byte float

Degrees Celsius

Read

 

 

 

 

 

147

Temperature status code

16-bit integer

 

Read

 

 

 

 

 

148-153

Temperature time record

Time

 

Read

 

 

 

 

 

160,161

Internal Temp

4-byte float

Degrees Celsius

Read

 

 

 

 

 

162

Internal Temp status code

16-bit integer

 

Read

 

 

 

 

 

163-168

Internal Temp time record

Time

 

Read

 

 

 

 

 

175,176

Analog channel 1

4-byte float

0-100 percent

Read

 

 

 

 

 

177

Analog channel 1 status code

16-bit integer

 

Read

 

 

 

 

 

178-183

Analog channel 1 time record

Time

 

Read

 

 

 

 

 

190,191

Analog channel 2

4-byte float

0-100 percent

Read

 

 

 

 

 

192

Analog channel 2 status code

16-bit integer

 

Read

 

 

 

 

 

193-198

Analog channel 2 time Record

Time

 

Read

 

 

 

 

 

205,206

Analog channel 3

4-byte float

0-100 percent

Read

 

 

 

 

 

207

Analog channel 3 status code

16-bit integer

 

Read

 

 

 

 

 

208-213

Analog channel 3 time record

Time

 

Read

 

 

 

 

 

220,221

Analog channel 4

4-byte float

0-100 percent

Read

 

 

 

 

 

222

Analog channel 4 status code

16-bit integer

 

Read

 

 

 

 

 

223-228

Analog channel 4 time record

Time

 

Read

 

 

 

 

 

235,236

Analog channel 5

4-byte float

0-100 percent

Read

 

 

 

 

 

237

Analog channel 5 status code

16-bit integer

 

Read

 

 

 

 

 

4-8

Image 50
Contents Field Wizard Page Foreword Page Table of Contents Field WizardAppendix B Accessories MaintenanceAppendix a Replacement Parts Appendix C Material Safety Data SheetsList of Tables Field Wizard Table of Contents Introduction Introduction Product DescriptionIdentifying Module Components Field Wizard Keypad and DisplayField Wizard Components Side View Field Wizard Components Bottom View 2101 Field Wizard Technical Specifications Safety Symbols and Hazard Alerts Technical ServicePreparation and Installation Unpacking InstructionsSafety Connecting To Flowlink Connecting the AC Adapter to the Field Wizard Connecting To a Site Flowlink Site Info ScreenUnlocking Unlocking the latch Connectors Field Wizard module unlock the latch Align the modules Data Retrieval Mode Main Display Operation Field Wizard KeypadNumeric Keys Power Key Enter Key Field Wizard Operation Flowlink Site Info Screen Bar Field Wizard Connected to a 2150 Site Data Retrieval Mode Site Time SynchronizationKEY Function Description Retrieve Retrieve Data Display Retrieve # Days Display Data Retrieval Log 10 Data Log Details Display 11 Field Wizard Site Setup Display 13 Site Setup Display Data Display Standalone Mode 17 Site Information DisplayPC Data Synchronization Mode 19 Standalone Information DisplayField Wizard Operation Modbus Protocol Establishing Configurations Configuration Example Direct Connection ShownGlossary of Terms Common Acronyms Modbus Ascii Address 1 Register Definitions Modbus Ascii Address 2-N+1 Register Definitions 117 Volume 1 status code Bit integer Read Measurement Parameters by Model Number Field Wizard Modbus Protocol Maintenance Kit MaintenanceMaintenance Overview Desiccant DesiccantHydrophobic Filter How to Obtain Service Field Wizard Maintenance Appendix a Replacement Parts Replacement PartsField Wizard Appendix a Replacement Parts Field Wizard Appendix a Replacement Parts Field Wizard Appendix a Replacement Parts Appendix B Accessories How to Order General AccessoriesField Wizard Appendix B Accessories Appendix C Material Safety Data Sheets OverviewMaterial Safety Data Sheet First Aid Measures Exposure Controls/Personal Protection Toxicological Information Other Information LVSOD\ Lulqj \SDG Page Declaration of Conformity Page Rmity Page Teledyne Isco One Year Limited Factory Service Warranty