2101 Field Wizard

Section 3 Operation

2100 Series Site Setup – If you press F2 NEXT and advance to a 2100 Series module, the Setup display shown in Figure 3-13 will appear.

Figure 3-13 Site Setup Display

The Site Setup display is used to calibrate sensors and run diag- nostics on sensors. Module sensor information, such as level and flow rate, is presented inside the display window. Use the Up and Down arrow keys to scroll within the window.

Measurement Calibration – Level sensors can be calibrated using the Field Wizard. Before you can calibrate, one valid reading must have been taken so there is a value to calibrate to.

On the Site Setup display, use F2 NEXT to display the module you want to calibrate, then press F3 CALIB. The Level Cali- bration window for the selected module will appear (Figure 3-14).

Figure 3-14 Level Calibration Display

The level currently being detected will be displayed. To change the level, use the numeric keys and then press Enter. You will be returned to the Site Setup display. The message “Updating” will appear in the Level field momentarily while the system updates its information.

To clear an entry, press F4 CLEAR. To cancel the process and return to the Site Setup display, press F6 CANCEL.

3-14

Image 38
Contents Field Wizard Page Foreword Page Table of Contents Field WizardAppendix B Accessories MaintenanceAppendix a Replacement Parts Appendix C Material Safety Data SheetsList of Tables Field Wizard Table of Contents Introduction Introduction Product DescriptionIdentifying Module Components Field Wizard Keypad and DisplayField Wizard Components Side View Field Wizard Components Bottom View 2101 Field Wizard Technical Specifications Safety Symbols and Hazard Alerts Technical ServicePreparation and Installation Unpacking InstructionsSafety Connecting To Flowlink Connecting the AC Adapter to the Field Wizard Connecting To a Site Flowlink Site Info ScreenUnlocking Unlocking the latch Connectors Field Wizard module unlock the latch Align the modules Data Retrieval Mode Main Display Operation Field Wizard KeypadNumeric Keys Power Key Enter Key Field Wizard Operation Flowlink Site Info Screen Bar Field Wizard Connected to a 2150 Site Data Retrieval Mode Site Time SynchronizationKEY Function Description Retrieve Retrieve Data Display Retrieve # Days Display Data Retrieval Log 10 Data Log Details Display 11 Field Wizard Site Setup Display 13 Site Setup Display Data Display Standalone Mode 17 Site Information DisplayPC Data Synchronization Mode 19 Standalone Information DisplayField Wizard Operation Modbus Protocol Establishing Configurations Configuration Example Direct Connection ShownGlossary of Terms Common Acronyms Modbus Ascii Address 1 Register Definitions Modbus Ascii Address 2-N+1 Register Definitions 117 Volume 1 status code Bit integer Read Measurement Parameters by Model Number Field Wizard Modbus Protocol Maintenance Kit MaintenanceMaintenance Overview Desiccant DesiccantHydrophobic Filter How to Obtain Service Field Wizard Maintenance Appendix a Replacement Parts Replacement PartsField Wizard Appendix a Replacement Parts Field Wizard Appendix a Replacement Parts Field Wizard Appendix a Replacement Parts Appendix B Accessories How to Order General AccessoriesField Wizard Appendix B Accessories Appendix C Material Safety Data Sheets OverviewMaterial Safety Data Sheet First Aid Measures Exposure Controls/Personal Protection Toxicological Information Other Information LVSOD\ Lulqj \SDG Page Declaration of Conformity Page Rmity Page Teledyne Isco One Year Limited Factory Service Warranty