Omega 3400ST, 3200ST manual Connecting The 6 Pin Mini Din Connector & FMA 3000C Cable

Page 10

c)Connecting The 6 Pin Mini Din Connector & FMA 3000C Cable

The two mating connectors should be pushed together and the pigtail leads wired as follows:

Connecting To The Integrated 6 Pin Connector Using A FMA 3000C Cable

The RED wire should be connected to the Positive of the power source.

The BLACK wire should be connected to the Negative (Ground) of the power source.

The ORANGE wire provides the signal output and should be connected to the positive terminal of the display, data acquisition system or voltmeter.

The BROWN wire is the signal negative (ground) and should be connected to the negative (Ground) terminal of the display, data acquisition system or voltmeter.

The YELLOW wire provides the input signal and should be connected to the positive terminal of the voltage source. The (0-5VDC) voltage control signal should be supplied from a low impedance source.

The GREEN wire is the input signal negative (ground) and should be connected to the negative (Ground) terminal of the voltage source.

The wire colors above describe the pigtail leads of the FMA 3000C cable assembly and may not correspond with the internal wiring of your flow sensor.

Caution: Avoid high voltage static discharges to the input signal connection. Do not short the input/output signal wires or allow them to contact the power wires at any time. DAMAGE WILL RESULT!

M-4271/0707, pg. 10 of 26

Image 10
Contents Page Page General Unpacking IntroductionProduct Overview and Principle of Operation Installation General Considerations4271/0707, pg Mounting the Flow Controller Tubing ConnectionsElectrical Connections Overview Connecting The 6 Pin Mini Din Connector Connecting The 6 Pin Mini Din Connector & FMA 3000C Cable Connections For The 9 Pin D Sub Connector Connections For The 15 D Sub Connector Using a 0-5VDC Input / Output Power Adapter Package Verification of Zero OperationWarm Up Flow ReadingsQ1 / Q2 = K1 / K2 Power Save Mode Changing The Flow Rate Set-Point FMA3400/3400ST Series OnlyRecalibration Zero AdjustmentsChanging the Calibration Gas FMA 3400/3400ST Series Only General Maintenance and Product CareReturning Units for Repair or Recalibration Specifications Series FMA FMA 3200ST FMA 3400STDimensions FMA 3400/3400ST Series 1/4 Stainless Fittings Shown Gas K Factors Gas ChemicalTroubleshooting Guide Symptom Possible Cause Method of CorrectionSymptom Possible Cause Method of Correction 4271/0707, pg 4271/0707, pg