Omega manual Specifications, Series FMA FMA 3200ST FMA 3400ST

Page 19

E.Specifications

 

Series

FMA 3200

 

 

FMA 3400

 

 

FMA 3200ST

 

 

 

FMA 3400ST

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accuracy

±1.5% of

 

 

±1.5% of

 

 

±1.5% of

 

±1.5% of

 

(including linearity)

Full Scale*

 

 

Full Scale*

 

 

Full Scale*

 

Full Scale*

 

 

 

 

 

Second and

 

 

 

 

 

 

Second and

 

 

 

 

 

third gases

 

 

 

 

 

 

third gases

 

 

 

 

 

±3.0% F.S.*

 

 

 

 

 

 

±3.0% F.S.*

 

 

 

 

 

 

 

 

 

 

Repeatability

 

 

 

±0.25% Full Scale*

 

 

 

 

 

 

 

 

Pressure Rating

150 psig (10.3 bar)

 

 

500 psig (34.5 bar)

 

 

 

 

 

 

 

 

 

Pressure Sensitivity

 

 

 

±0.02% Full Scale* per psi (per 69 mbar)

 

 

 

 

 

 

 

 

 

Temperature Rating

 

 

 

Operating Range: 5 to 55ºC

 

 

 

Recommended Range (for best performance) : 10 to 40ºC

 

 

 

 

 

Storage Range: 0 to 70ºC

 

 

 

 

 

 

 

 

 

Temperature

 

 

 

±0.15% F.S.* or less per ºC

 

 

Sensitivity

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Valve

 

 

 

Normally Closed

 

 

 

 

 

 

Positive Shut-off up to 60psig (4 bar)

 

 

 

 

 

 

 

 

 

Body Leak Integrity

 

 

 

1x10-7sccs of He

 

 

 

 

 

 

 

 

Wetted Materials

Aluminum

 

 

303 Stainless Steel

 

 

304 Stainless Steel

 

 

304 Stainless Steel

 

 

316 Stainless Steel

 

 

316 Stainless Steel

 

 

 

 

 

 

 

 

Epoxy

 

 

 

 

 

 

 

 

 

 

 

O-Ring Material

 

 

 

 

Viton®

 

 

 

 

 

 

 

 

Fitting Material

 

 

Choose from acetal, brass, or stainless steel

 

 

 

 

 

 

 

 

 

Recommended

 

 

 

20 microns or less

 

 

Filtration

 

 

 

Optional inline filters available

 

 

 

 

 

 

Compatible gases

 

Clean, dry gases compatible with wetted materials

 

 

 

 

 

 

 

 

Output Signal

 

 

 

0-5VDC, Impedance greater than 2.5 KΩ

 

 

 

 

 

 

 

 

 

External Set-point

 

 

 

0-5VDC, Integrated 2MΩ load

 

 

Signal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Internal Set-point

N/A

 

 

Front Panel

 

 

N/A

 

Front Panel

 

Signal

 

 

 

Adjustment

 

 

 

 

 

 

Adjustment

 

 

 

 

 

 

 

 

 

 

Warm-Up Time

 

 

 

Less than 5 minutes

 

 

 

 

 

 

 

 

 

 

 

 

Integrated Display

N/A

 

 

3½ digit

 

 

N/A

 

3½ digit

 

 

 

 

 

 

 

 

 

 

Typical Power

 

 

Standard: 12 VDC @ 250 mA (12.5-15 VDC)

 

 

Consumption

 

 

“E” Suffix: 24 VDC @ 130 mA (22-25 VDC)

 

 

 

 

 

 

 

 

Peak Power

 

 

Standard: 12 VDC @ 400 mA (12.5-15 VDC)

 

 

Consumption

 

 

“E” Suffix: 24 VDC @ 260 mA (22-25 VDC)

 

 

 

 

 

 

 

Electrical

 

Integrated 36” (92 mm) cable, terminated with:

 

 

Connections

 

 

Standard: 6-pin Mini-DIN male (PS/2 Style)

 

 

 

 

 

 

D1 Option: 9-pin D-Sub male

 

 

 

 

 

 

D2 Option: 15-pin D-Sub male

 

 

 

 

 

 

 

 

 

Certifications

 

 

 

CE Approved

 

 

 

 

 

 

89/336/EEC (EN 55011 & EN 50082-1)

 

 

 

 

 

 

73/23/EEC Low Voltage Directive

 

 

 

 

 

 

 

 

 

 

 

 

 

 

*Specifications from 10-100% of rated flow. Linearity is best fit straight line. All calibrations performed with air unless otherwise stated on calibration certificate.

M-4271/0707, pg. 19 of 26

Image 19
Contents Page Page General Unpacking IntroductionProduct Overview and Principle of Operation General Considerations Installation4271/0707, pg Tubing Connections Mounting the Flow ControllerElectrical Connections Overview Connecting The 6 Pin Mini Din Connector Connecting The 6 Pin Mini Din Connector & FMA 3000C Cable Connections For The 9 Pin D Sub Connector Connections For The 15 D Sub Connector Using a 0-5VDC Input / Output Power Adapter Package Flow Readings OperationWarm Up Verification of ZeroQ1 / Q2 = K1 / K2 Changing The Flow Rate Set-Point FMA3400/3400ST Series Only Power Save ModeRecalibration Zero AdjustmentsChanging the Calibration Gas FMA 3400/3400ST Series Only General Maintenance and Product CareReturning Units for Repair or Recalibration Series FMA FMA 3200ST FMA 3400ST SpecificationsDimensions FMA 3400/3400ST Series 1/4 Stainless Fittings Shown Gas Chemical Gas K FactorsSymptom Possible Cause Method of Correction Troubleshooting GuideSymptom Possible Cause Method of Correction 4271/0707, pg 4271/0707, pg