Omega 3200ST, 3400ST manual Electrical Connections Overview

Page 8

4.Electrical Connections

Caution: Incorrect wiring may cause severe damage to the unit. Applying an AC voltage (115VAC or 230VAC) directly to the unit will cause damage. Read the following instructions carefully before making any connections.

a)Overview

The FMA 3200/3400 Series provides a 0-5VDC analog output proportional to the flow rate. This output may be connected to a display, data acquisition system or voltmeter with an impedance of greater than 2.5 kΩ (kilo ohms).

The flow controller needs to be supplied with a 0-5 VDC set point signal to enable control. On the FMA 3400/3400ST Series this may be generated internally by altering the set-point potentiometer on the front panel of the unit.

A stable D.C. power supply is required to operate the unit. The voltage and current requirements depend on the configuration of the unit. Full details may be found in the Specification section of this manual.

Connecting wires should be as short as possible to avoid voltage drops. Twisted conductor cable should be used if the length of the wiring is to be longer than 2 meters.

Units are supplied with either a 6 pin mini DIN type connector (requires mating cable assembly), a 9 Pin D Sub connector or 15 Pin D Sub connector.

Caution: Cutting off the integrated connectors on the unit IS NOT RECOMMENDED and will void the product warranty. Mating cables should be ordered along with each unit.

Electrical connections to the units are made as detailed in the following sections.

M-4271/0707, pg. 8 of 26

Image 8
Contents Page Page General Product Overview and Principle of Operation IntroductionUnpacking Installation General Considerations4271/0707, pg Mounting the Flow Controller Tubing ConnectionsElectrical Connections Overview Connecting The 6 Pin Mini Din Connector Connecting The 6 Pin Mini Din Connector & FMA 3000C Cable Connections For The 9 Pin D Sub Connector Connections For The 15 D Sub Connector Using a 0-5VDC Input / Output Power Adapter Package Operation Warm UpVerification of Zero Flow ReadingsQ1 / Q2 = K1 / K2 Power Save Mode Changing The Flow Rate Set-Point FMA3400/3400ST Series OnlyChanging the Calibration Gas FMA 3400/3400ST Series Only Zero AdjustmentsRecalibration Returning Units for Repair or Recalibration Maintenance and Product CareGeneral Specifications Series FMA FMA 3200ST FMA 3400STDimensions FMA 3400/3400ST Series 1/4 Stainless Fittings Shown Gas K Factors Gas ChemicalTroubleshooting Guide Symptom Possible Cause Method of CorrectionSymptom Possible Cause Method of Correction 4271/0707, pg 4271/0707, pg