A&D AD-4212A/B DP Dump print format 5if type, KF format 5if type, MT format 5if type

Page 39

DP (Dump print) format

5if type 1

This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B is used, set the printer to MODE 3.

This format consists of sixteen characters excluding the terminator.

A header of two characters indicates the balance condition. No overload header is used.

The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not zero or overloaded.

The unit, consisting of three characters, follows the data.

W T

+0

. 1 2 7 8

g

CR LF

Header Data

WT Stable header U S Unstable header

Q T

Unit Terminator

Stable header of counting mode

KF format

5if type 2

This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only communicate using this format.

This format consists of fourteen characters excluding the terminator. This format has no header characters.

The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not zero or overloaded.

This format outputs the unit only for a stable value.

Data

 

 

 

 

 

 

 

Unit Terminator

 

 

 

 

 

 

 

 

 

 

 

 

 

Stable value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Unstable value

 

 

 

 

 

 

 

 

 

 

 

 

 

MT format

5if type 3

A header of two characters indicates the balance condition.

The polarity sign is used only for negative data.

The weighing data uses spaces in place of the leading zeros.

The character length of this format changes dependent upon the unit

S

0 .

1

2

7

8

 

 

g

 

 

 

 

 

 

 

CR LF

Header

Data

Unit Terminator

S Stable header

S D Unstable header S I Overload header

NU (numerical) format

5if type 4

This format outputs only numerical data.

This format consists of nine characters for the AD-4212A series balance or ten characters for the AD-4212B series balance, excluding the terminator.

The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is used.

+ 0 0 0 . 1 2 7 8 CR LF

(AD-4212A series balance)

Data

+0 0 0 0 . 1 2 7 Data

Terminator

8 CR LF (AD-4212B series balance)

Terminator

37

Image 39
Contents 1WMPD4001448C 212 BPage Contents ID Number and GLP Report TERMS/INDEX Introduction FeaturesWindows is a registered trademark of Microsoft Corporation ComplianceCompliance with FCC Rules Compliance with EMC DirectivesUnpacking the Balance UnpackingRear of the display unit AC adapter ID label AccessoriesThat it fits over the dust guard Installing the BalanceHow to install the breeze break Place the breeze break on the weighing pan soRear of the weighing unit How to disconnect the cable from the weighing unitBefore Use PrecautionsWhen Building into a System Lighter heavier than the true weight. This Power Supply During UseAfter Use Stabilization indicator Display Symbols and KEY OperationDisplay symbols Processing indicatorWeighing Units UnitsProgrammable-unit Capacity Minimum Display Minimum CapacityChanging the Unit Weighing Basic Operation Gram ModeSmart Range Function Precision range/standard range value01 mg 0.1 mg Standard range Sensitive value Changing the Weighing SpeedOperation CondDisplay CalibrationCalibration Calibration testCalibration From previous Return to CC Usable weights is shown on page 21, pressCalibration Test Key until CCoutFrom previous For details, refer to 20. Extended Functions Function Switch and InitializationSwitches Permit or InhibitBe sure to calibrate the balance after initialization Initializing the BalanceExample Function TableStructure and Sequence of the Function Table Display and KeysDetails of the Function Table Only the AD-4212B series Output SettingData output mode Cpout Cp-tTime/Date output Auto print polarityData memory Interval timeBaud rate TimeoutAK, Error code ID number settingHold function Hold Animal weighing mode, AD-4212A only Description of the Class Environment, DisplayCondition Cond Stability band width 5t-bDecimal point pnt Capacity indicator g5iZero tracking trc Display refresh rate 5pdTare at start p-tr Description of the Item Data output modeKey mode Stream modeDescription of the Item Data format Key mode BKey mode C Interval memory modeNU numerical format 5if type DP Dump print format 5if typeKF format 5if type MT format 5if typeDate dout 5-td 2 or Description of the Data Format Added to the Weighing DataID number Dout 5-id Data number dout d-no20h Positive errorNegative error Space Data Format ExamplesSetting the time with part of the digits blinking Clock and Calendar FunctionConfirming the time Set the date using the following keys Confirming the dateQuitting the operation Comparator Function Three-level comparison resultsFive-level comparison results Indicates a space 20h Setting the upper/lower limit valuesWhen three-level comparison is selected When five-level comparison is selectedConfirming the upper/lower limit values Adding the comparison results AD-4212A only Setting the ID Number ID Number and GLP ReportDisplay character set Set the following parameters to output the report When the setting is infoGLP Report Calibration reportCalibration test report using an external weight Title block and end block Weighing dataEnd block Counting Mode PC Selecting the counting modeStoring a sample unit mass Counting operation Counting mode using the Acai functionSelecting the percent mode Storing the 100% reference massReading the percentage Key until SClr no Balance displays End Returns to the weighing modeData Memory Releasing ClrStable, the balance stores the weighing data Memory for Weighing DataStoring the weighing data Key is pressed while the displayed value isSetting the function table Press the CAL key to return to the weighing modeRecalling the memory data Deleting all memory data at one time Transmitting all memory data at one timeDeleting the memory data Memory for Calibration and Calibration Test DataStoring the calibration and calibration test data Transmitting the memory dataKey To increase the unit mass number by one Memory for Unit Mass in the Counting ModeKey To cancel the operation and go to step Recalling the unit massMemory for Upper/lower limit values Inputting the upper/lower limit values 3-level comparison Switching the upper/lower limit values 3-level comparison Selecting the programmable-unit Setting the coefficientUsing the function Confirming the coefficientRS-232C Sub 25 pin numbers Sub 25 pin assignmentsUse a shielded cable Use a shieldedCable External contact inputConnection to Peripheral Equipment Mode PrtWhen multiple lines are to be printed, set the parameter to AD-8121B settingsRsKey Using Windows Communication Tools WinCTConnection to a Computer RsComUsing the WinCT software, the balance can do the following Commands to control the balance CommandsCommand List Commands to query weighing dataAcknowledge Code and Error Codes Settings Related to RS-232C Control Using CTS and RTSCable Connector Pin Nos. and SpecificationsPin assignments and I/O logic OP-01 installation procedure Timing chartOn, T off 100 m sec or longer Extended Functions AD-4212A only Page Averaging range f1-b and averaging time f1-t Description of Averaging range and Averaging timeChecking the Balance Performance and Environment MaintenanceTroubleshooting Check the balance performance using the self-check functionOperating environment Weighing methodSample and container Error Codes EC, E00 Communications error Unit mass errorClock battery error Memory type errorAsking For Repair Specifications Counting Minimum unit mass Mode Calibration weight providedDisplay Dimensions 237 W × 150 D × 155 H mm Mass Power receptacle type Power consumptionUsing the pan support Designing a Special Weighing PANPan support can be removed Using the pan boss Shock absorber specificationsPan boss can not be removed Mass of the special weighing pan AD-4212A series balance AD-4212A-100AD-4212A-200 AD-4212A-1000 AD-4212A-600AD-4212B-101/102 Mass of the special weighing pan AD-4212B series balanceAD-4212B-301 AD-4212B-201Installing in a panel Installing the Display UnitAttachment Procedure Attaching the Mounting FixturesSecure the weighing unit from above Weighing unit External DimensionsDisplay unit Same for all the modelsWeighing unit with the mounting fixtures attached AD-1684 Electrostatic field meter OptionsAD-8121B Printer AD-1683 DC static eliminatorOP-20 Metal leveling foot OP-08 Ethernet interfaceTERMS/INDEX TermsIndex Cp HH BepCl adj Cp fncRng G5iPnt Off5pd 5t-bTrc Memo

AD-4212A-100/200/600/1000, AD-4212A/B, AD-4212B-101/102/201/301 specifications

The A&D AD-4212 series, including models AD-4212B-101, AD-4212B-102, AD-4212B-201, AD-4212B-301, AD-4212A/B, and AD-4212A-100/200/600/1000, represents a significant advancement in precision weighing technology. A&D Company, a leader in the field of measurement and weighing solutions, has equipped these models with various features that cater to diverse industrial needs.

One of the standout characteristics of the AD-4212 series is its high precision load cell technology. This technology enables accurate measurements with a resolution that can reach up to 0.01 grams, making it suitable for industries where precision is paramount, such as pharmaceuticals and laboratories. Additionally, the series boasts an impressive overload protection feature, allowing it to handle weights significantly above its rated capacity without sustaining damage.

Durability and user-friendliness are also top priorities for the AD-4212 models. The devices are constructed using high-quality materials that ensure longevity, even in demanding environments. The intuitive user interface includes a large, easy-to-read LCD display, which enhances usability by allowing operators to view measurements and settings effortlessly.

Another essential feature of the AD-4212 series is its versatility in connectivity. These models support multiple communication protocols, including USB and RS-232, enabling seamless integration with computers and other devices for data output and printing. This connectivity capability facilitates streamlined operations in various settings, allowing for efficient data management and record-keeping.

User customization is also a significant aspect of these models. The AD-4212 series allows users to calibrate the devices according to their specific requirements, ensuring that they can tailor the performance to suit varied applications. Moreover, the models come with built-in functions for tare, hold, counting, and statistical calculations, making them extremely flexible for different weighing scenarios.

Energy efficiency is another noteworthy consideration in the design of the AD-4212 series. These models often feature automatic shut-off functions that help conserve energy, extending battery life and reducing operating costs. The combination of these technologies and characteristics makes the A&D AD-4212 series an exemplary choice for professionals seeking reliable and accurate weighing solutions across various industries. Overall, the AD-4212 models exemplify A&D's commitment to quality, precision, and user satisfaction.