Page 80

When the balance is built into a system, remove the balance from the system. Place it on a solid table. Install the breeze break and perform checking.

When the balance proper performance is confirmed, refer to page 10 to set up the installation site.

Checking that the operating environment or weighing method is proper

Operating environment

Is the weighing table solid enough

Is the balance level? Refer to “3-1 Before Use”.

Is the operating environment free from vibration and drafts? Has the stainless steel breeze break been installed?

Is there a strong electrical or magnetic noise source such as a motor near the balance? Is there a heat source near the balance?

Weighing method

Does the weighing pan touch the breeze break or anything? Is the weighing pan installed correctly?

Is the RE-ZERO key pressed before placing a sample on the weighing pan? Or, is RE-ZERO performed using external contact input or an RS-232C command?

Is the sample placed in the center of the weighing pan?

Has the balance been warmed up for the appropriate duration before weighing? AD-4212A series balance: 30 minutes or more

AD-4212B series balance: one hour or more

Are the leveling feet of the weighing unit placed flat to the installation surface?

If not, the weight value will be unstable or the specified repeatability can not be obtained. Improve the installation condition, by securing the weighing unit or reducing the vibration that is conveyed to the weighing unit.

Sample and container

Has the sample absorbed or lost moisture due to the ambient conditions such as temperature or humidity?

Has the temperature of the container been allowed to equalize to the ambient temperature? Refer to “3-2 When Building into a System”.

Is the sample charged with static electricity? Refer to “3-2 When Building into a System”.

Is the sample of magnetic material such as iron? Use much care when weighing magnetic materials. Refer to “3-2 When Building into a System”.

78

Image 80
Contents 212 B 1WMPD4001448CPage Contents ID Number and GLP Report TERMS/INDEX Features IntroductionCompliance Compliance with FCC RulesCompliance with EMC Directives Windows is a registered trademark of Microsoft CorporationRear of the display unit Unpacking the BalanceUnpacking Accessories AC adapter ID labelInstalling the Balance How to install the breeze breakPlace the breeze break on the weighing pan so That it fits over the dust guardHow to disconnect the cable from the weighing unit Rear of the weighing unitPrecautions Before UseWhen Building into a System Lighter heavier than the true weight. This After Use Power SupplyDuring Use Display Symbols and KEY Operation Display symbolsProcessing indicator Stabilization indicatorProgrammable-unit Weighing UnitsUnits Minimum Capacity Capacity Minimum DisplayChanging the Unit Basic Operation Gram Mode Weighing01 mg 0.1 mg Standard range Smart Range FunctionPrecision range/standard range value Changing the Weighing Speed OperationCond Sensitive valueCalibration CalibrationCalibration test DisplayCalibration From previous Usable weights is shown on page 21, press Calibration TestKey until CCout Return to CCFrom previous Function Switch and Initialization SwitchesPermit or Inhibit For details, refer to 20. Extended FunctionsInitializing the Balance Be sure to calibrate the balance after initializationFunction Table Structure and Sequence of the Function TableDisplay and Keys ExampleDetails of the Function Table Output Setting Data output modeCpout Cp-t Only the AD-4212B seriesAuto print polarity Data memoryInterval time Time/Date outputTimeout AK, Error codeID number setting Baud rateDescription of the Class Environment, Display Condition CondStability band width 5t-b Hold function Hold Animal weighing mode, AD-4212A onlyCapacity indicator g5i Zero tracking trcDisplay refresh rate 5pd Decimal point pntDescription of the Item Data output mode Key modeStream mode Tare at start p-trKey mode B Key mode CInterval memory mode Description of the Item Data formatDP Dump print format 5if type KF format 5if typeMT format 5if type NU numerical format 5if typeDescription of the Data Format Added to the Weighing Data ID number Dout 5-idData number dout d-no Date dout 5-td 2 orPositive error Negative error SpaceData Format Examples 20hConfirming the time Setting the time with part of the digits blinkingClock and Calendar Function Quitting the operation Set the date using the following keysConfirming the date Five-level comparison results Comparator FunctionThree-level comparison results Setting the upper/lower limit values Indicates a space 20hConfirming the upper/lower limit values When three-level comparison is selectedWhen five-level comparison is selected Adding the comparison results AD-4212A only Display character set Setting the ID NumberID Number and GLP Report When the setting is info GLP ReportCalibration report Set the following parameters to output the reportCalibration test report using an external weight End block Title block and end blockWeighing data Storing a sample unit mass Counting Mode PCSelecting the counting mode Counting mode using the Acai function Counting operationReading the percentage Selecting the percent modeStoring the 100% reference mass Balance displays End Returns to the weighing mode Data MemoryReleasing Clr Key until SClr noMemory for Weighing Data Storing the weighing dataKey is pressed while the displayed value is Stable, the balance stores the weighing dataRecalling the memory data Setting the function tablePress the CAL key to return to the weighing mode Transmitting all memory data at one time Deleting all memory data at one timeMemory for Calibration and Calibration Test Data Storing the calibration and calibration test dataTransmitting the memory data Deleting the memory dataMemory for Unit Mass in the Counting Mode Key To cancel the operation and go to stepRecalling the unit mass Key To increase the unit mass number by oneMemory for Upper/lower limit values Inputting the upper/lower limit values 3-level comparison Switching the upper/lower limit values 3-level comparison Setting the coefficient Using the functionConfirming the coefficient Selecting the programmable-unitSub 25 pin numbers Sub 25 pin assignments RS-232CUse a shielded CableExternal contact input Use a shielded cableMode Prt When multiple lines are to be printed, set the parameter toAD-8121B settings Connection to Peripheral EquipmentUsing Windows Communication Tools WinCT Connection to a ComputerRsCom RsKeyUsing the WinCT software, the balance can do the following Commands Command ListCommands to query weighing data Commands to control the balanceAcknowledge Code and Error Codes Control Using CTS and RTS Settings Related to RS-232CConnector Pin Nos. and Specifications CablePin assignments and I/O logic On, T off 100 m sec or longer OP-01 installation procedureTiming chart Extended Functions AD-4212A only Page Description of Averaging range and Averaging time Averaging range f1-b and averaging time f1-tMaintenance TroubleshootingCheck the balance performance using the self-check function Checking the Balance Performance and EnvironmentSample and container Operating environmentWeighing method Error Codes Unit mass error Clock battery errorMemory type error EC, E00 Communications errorAsking For Repair Specifications Calibration weight provided Display Dimensions 237 W × 150 D × 155 H mm MassPower receptacle type Power consumption Counting Minimum unit mass ModePan support can be removed Using the pan supportDesigning a Special Weighing PAN Pan boss can not be removed Using the pan bossShock absorber specifications AD-4212A-200 Mass of the special weighing pan AD-4212A series balanceAD-4212A-100 AD-4212A-600 AD-4212A-1000Mass of the special weighing pan AD-4212B series balance AD-4212B-101/102AD-4212B-201 AD-4212B-301Installing the Display Unit Installing in a panelSecure the weighing unit from above Attachment ProcedureAttaching the Mounting Fixtures External Dimensions Display unitSame for all the models Weighing unitWeighing unit with the mounting fixtures attached Options AD-8121B PrinterAD-1683 DC static eliminator AD-1684 Electrostatic field meterOP-08 Ethernet interface OP-20 Metal leveling footIndex TERMS/INDEXTerms Bep Cl adjCp fnc Cp HHG5i PntOff RngTrc 5pd5t-b Memo

AD-4212A-100/200/600/1000, AD-4212A/B, AD-4212B-101/102/201/301 specifications

The A&D AD-4212 series, including models AD-4212B-101, AD-4212B-102, AD-4212B-201, AD-4212B-301, AD-4212A/B, and AD-4212A-100/200/600/1000, represents a significant advancement in precision weighing technology. A&D Company, a leader in the field of measurement and weighing solutions, has equipped these models with various features that cater to diverse industrial needs.

One of the standout characteristics of the AD-4212 series is its high precision load cell technology. This technology enables accurate measurements with a resolution that can reach up to 0.01 grams, making it suitable for industries where precision is paramount, such as pharmaceuticals and laboratories. Additionally, the series boasts an impressive overload protection feature, allowing it to handle weights significantly above its rated capacity without sustaining damage.

Durability and user-friendliness are also top priorities for the AD-4212 models. The devices are constructed using high-quality materials that ensure longevity, even in demanding environments. The intuitive user interface includes a large, easy-to-read LCD display, which enhances usability by allowing operators to view measurements and settings effortlessly.

Another essential feature of the AD-4212 series is its versatility in connectivity. These models support multiple communication protocols, including USB and RS-232, enabling seamless integration with computers and other devices for data output and printing. This connectivity capability facilitates streamlined operations in various settings, allowing for efficient data management and record-keeping.

User customization is also a significant aspect of these models. The AD-4212 series allows users to calibrate the devices according to their specific requirements, ensuring that they can tailor the performance to suit varied applications. Moreover, the models come with built-in functions for tare, hold, counting, and statistical calculations, making them extremely flexible for different weighing scenarios.

Energy efficiency is another noteworthy consideration in the design of the AD-4212 series. These models often feature automatic shut-off functions that help conserve energy, extending battery life and reducing operating costs. The combination of these technologies and characteristics makes the A&D AD-4212 series an exemplary choice for professionals seeking reliable and accurate weighing solutions across various industries. Overall, the AD-4212 models exemplify A&D's commitment to quality, precision, and user satisfaction.