A&D AD-4212A/B instruction manual Setting the upper/lower limit values, Indicates a space 20h

Page 45

Five-level comparison

Comparison

 

 

 

 

 

Display

 

Contact output

 

Buzzer

results

 

 

 

 

 

H H

HI

OK

LO

LL

 

 

 

 

 

 

 

H H

 

 

 

 

 

blinking

short

open

open

open

open

 

HH

 

 

 

 

 

 

 

 

 

 

 

 

 

HI

 

 

HI illuminated

open

short

open

open

open

 

OK

 

 

 

 

illuminated

open

open

short

open

open

 

 

OK

 

 

 

 

 

 

 

 

 

 

 

 

LO

 

 

 

illuminated

open

open

open

short

open

 

 

LO

 

 

 

 

 

 

 

 

 

 

LL

 

LL blinking

open

open

open

open

short

 

*Set the upper limit value equal to or higher than the lower limit value.

Set the secondary upper limit value equal to or higher than the upper limit value. Set the secondary lower limit value equal to or lower than the lower limit value.

Setting the upper/lower limit values

When five-level comparison is to be performed, using the AD-4212B series balance, set the “Comparator output setting ( Cp-t )” parameter to “1”.

1Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.

2Press the SAMPLE key several times to display the value name to be set.

CpPHH

CpPHi

CpPlo

CpPll

secondary upper limit value

upper limit value

lower limit value

secondary lower limit value

3Press the PRINT key. The value currently set is displayed with all of the digits blinking.

4When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 5.

When the current setting is to be changed, press the RE-ZERO key. Change the setting using the following keys.

SAMPLE

key

To select the digit to change the value.

 

 

 

key

To change the value of the digit selected.

RE-ZERO

 

 

key

To switch the polarity.

MODE

 

 

key

To store the new setting and go to step 5.

PRINT

 

key

To cancel the new setting and go to step 5.

CAL

5 Repeat steps 2-4 to set values as necessary.

Note

The upper/lower limit values can be set by external commands using the RS-232C serial interface.

Command example

H H :+100.00000

 

 

 

 

g (to set secondary upper limit value)

 

 

 

H I :+080.00000

 

 

 

 

g (to set upper limit value)

 

 

 

 

 

 

L O :+060.00000

 

 

 

 

 

g

(to set lower limit value)

 

LL :+040.00000

 

 

 

 

 

g

(to set secondary lower limit value)

 

 

 

 

 

 

 

 

 

 

 

 

indicates a space (20h)

 

 

 

 

 

 

43

Image 45
Contents 1WMPD4001448C 212 BPage Contents ID Number and GLP Report TERMS/INDEX Introduction FeaturesCompliance with FCC Rules ComplianceCompliance with EMC Directives Windows is a registered trademark of Microsoft CorporationUnpacking the Balance UnpackingRear of the display unit AC adapter ID label AccessoriesHow to install the breeze break Installing the BalancePlace the breeze break on the weighing pan so That it fits over the dust guardRear of the weighing unit How to disconnect the cable from the weighing unitBefore Use PrecautionsWhen Building into a System Lighter heavier than the true weight. This Power Supply During UseAfter Use Display symbols Display Symbols and KEY OperationProcessing indicator Stabilization indicatorWeighing Units UnitsProgrammable-unit Capacity Minimum Display Minimum CapacityChanging the Unit Weighing Basic Operation Gram ModeSmart Range Function Precision range/standard range value01 mg 0.1 mg Standard range Operation Changing the Weighing SpeedCond Sensitive valueCalibration CalibrationCalibration test DisplayCalibration From previous Calibration Test Usable weights is shown on page 21, pressKey until CCout Return to CCFrom previous Switches Function Switch and InitializationPermit or Inhibit For details, refer to 20. Extended FunctionsBe sure to calibrate the balance after initialization Initializing the BalanceStructure and Sequence of the Function Table Function TableDisplay and Keys ExampleDetails of the Function Table Data output mode Output SettingCpout Cp-t Only the AD-4212B seriesData memory Auto print polarityInterval time Time/Date outputAK, Error code TimeoutID number setting Baud rateCondition Cond Description of the Class Environment, DisplayStability band width 5t-b Hold function Hold Animal weighing mode, AD-4212A onlyZero tracking trc Capacity indicator g5iDisplay refresh rate 5pd Decimal point pntKey mode Description of the Item Data output modeStream mode Tare at start p-trKey mode C Key mode BInterval memory mode Description of the Item Data formatKF format 5if type DP Dump print format 5if typeMT format 5if type NU numerical format 5if typeID number Dout 5-id Description of the Data Format Added to the Weighing DataData number dout d-no Date dout 5-td 2 orNegative error Space Positive errorData Format Examples 20hSetting the time with part of the digits blinking Clock and Calendar FunctionConfirming the time Set the date using the following keys Confirming the dateQuitting the operation Comparator Function Three-level comparison resultsFive-level comparison results Indicates a space 20h Setting the upper/lower limit valuesWhen three-level comparison is selected When five-level comparison is selectedConfirming the upper/lower limit values Adding the comparison results AD-4212A only Setting the ID Number ID Number and GLP ReportDisplay character set GLP Report When the setting is infoCalibration report Set the following parameters to output the reportCalibration test report using an external weight Title block and end block Weighing dataEnd block Counting Mode PC Selecting the counting modeStoring a sample unit mass Counting operation Counting mode using the Acai functionSelecting the percent mode Storing the 100% reference massReading the percentage Data Memory Balance displays End Returns to the weighing modeReleasing Clr Key until SClr noStoring the weighing data Memory for Weighing DataKey is pressed while the displayed value is Stable, the balance stores the weighing dataSetting the function table Press the CAL key to return to the weighing modeRecalling the memory data Deleting all memory data at one time Transmitting all memory data at one timeStoring the calibration and calibration test data Memory for Calibration and Calibration Test DataTransmitting the memory data Deleting the memory dataKey To cancel the operation and go to step Memory for Unit Mass in the Counting ModeRecalling the unit mass Key To increase the unit mass number by oneMemory for Upper/lower limit values Inputting the upper/lower limit values 3-level comparison Switching the upper/lower limit values 3-level comparison Using the function Setting the coefficientConfirming the coefficient Selecting the programmable-unitRS-232C Sub 25 pin numbers Sub 25 pin assignmentsCable Use a shieldedExternal contact input Use a shielded cableWhen multiple lines are to be printed, set the parameter to Mode PrtAD-8121B settings Connection to Peripheral EquipmentConnection to a Computer Using Windows Communication Tools WinCTRsCom RsKeyUsing the WinCT software, the balance can do the following Command List CommandsCommands to query weighing data Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Control Using CTS and RTSCable Connector Pin Nos. and SpecificationsPin assignments and I/O logic OP-01 installation procedure Timing chartOn, T off 100 m sec or longer Extended Functions AD-4212A only Page Averaging range f1-b and averaging time f1-t Description of Averaging range and Averaging timeTroubleshooting MaintenanceCheck the balance performance using the self-check function Checking the Balance Performance and EnvironmentOperating environment Weighing methodSample and container Error Codes Clock battery error Unit mass errorMemory type error EC, E00 Communications errorAsking For Repair Specifications Display Dimensions 237 W × 150 D × 155 H mm Mass Calibration weight providedPower receptacle type Power consumption Counting Minimum unit mass ModeUsing the pan support Designing a Special Weighing PANPan support can be removed Using the pan boss Shock absorber specificationsPan boss can not be removed Mass of the special weighing pan AD-4212A series balance AD-4212A-100AD-4212A-200 AD-4212A-1000 AD-4212A-600AD-4212B-101/102 Mass of the special weighing pan AD-4212B series balanceAD-4212B-301 AD-4212B-201Installing in a panel Installing the Display UnitAttachment Procedure Attaching the Mounting FixturesSecure the weighing unit from above Display unit External DimensionsSame for all the models Weighing unitWeighing unit with the mounting fixtures attached AD-8121B Printer OptionsAD-1683 DC static eliminator AD-1684 Electrostatic field meterOP-20 Metal leveling foot OP-08 Ethernet interfaceTERMS/INDEX TermsIndex Cl adj BepCp fnc Cp HHPnt G5iOff Rng5pd 5t-bTrc Memo

AD-4212A-100/200/600/1000, AD-4212A/B, AD-4212B-101/102/201/301 specifications

The A&D AD-4212 series, including models AD-4212B-101, AD-4212B-102, AD-4212B-201, AD-4212B-301, AD-4212A/B, and AD-4212A-100/200/600/1000, represents a significant advancement in precision weighing technology. A&D Company, a leader in the field of measurement and weighing solutions, has equipped these models with various features that cater to diverse industrial needs.

One of the standout characteristics of the AD-4212 series is its high precision load cell technology. This technology enables accurate measurements with a resolution that can reach up to 0.01 grams, making it suitable for industries where precision is paramount, such as pharmaceuticals and laboratories. Additionally, the series boasts an impressive overload protection feature, allowing it to handle weights significantly above its rated capacity without sustaining damage.

Durability and user-friendliness are also top priorities for the AD-4212 models. The devices are constructed using high-quality materials that ensure longevity, even in demanding environments. The intuitive user interface includes a large, easy-to-read LCD display, which enhances usability by allowing operators to view measurements and settings effortlessly.

Another essential feature of the AD-4212 series is its versatility in connectivity. These models support multiple communication protocols, including USB and RS-232, enabling seamless integration with computers and other devices for data output and printing. This connectivity capability facilitates streamlined operations in various settings, allowing for efficient data management and record-keeping.

User customization is also a significant aspect of these models. The AD-4212 series allows users to calibrate the devices according to their specific requirements, ensuring that they can tailor the performance to suit varied applications. Moreover, the models come with built-in functions for tare, hold, counting, and statistical calculations, making them extremely flexible for different weighing scenarios.

Energy efficiency is another noteworthy consideration in the design of the AD-4212 series. These models often feature automatic shut-off functions that help conserve energy, extending battery life and reducing operating costs. The combination of these technologies and characteristics makes the A&D AD-4212 series an exemplary choice for professionals seeking reliable and accurate weighing solutions across various industries. Overall, the AD-4212 models exemplify A&D's commitment to quality, precision, and user satisfaction.