Agilent Technologies 66lxxA Init Initcont 1 Initcont on, Listcoun Listcoun INF, Listcurr

Page 36

Command Syntax

INITiate[:IMMediate]

 

 

INITiate:CONTinuous <bool>

Parameters

For INIT[:IMM]

(None)

 

For INIT:CONT

0 1 OFF ON

*RST Value

OFF

 

Examples

INIT INIT:CONT 1 INIT:CONT ON

Query Syntax

For INIT[:IMM]

(None)

 

For INIT:CONT

INITiate:CONTinuous?

Returned Parameters

0 1

 

Related Commands

ABOR CURR:TRIG

TRIG *TRG VOLT:TRIG

List Subsystem

This subsystem controls the generation of parameter lists that sequence the power module output through values of voltage and current. Two subsystem commands specify lists of output voltages (LIST:VOLT), and currents (LIST:CURR). A count command (LIST:COUN) determines how many times the power module sequences through a list before that list is completed. A dwell command (LIST:DWEL) specifies the time interval that each value (point) of a list is to remain in effect. A step command (LIST:STEP) determines if a trigger causes a list to advance only to its next point or to sequence through all of its points.

Each list can have from 1 to 20 points. Normally, voltage, current, and dwell lists must have the same number of points, or an error is generated when the first list point is triggered. The exception is a list consisting of only one point. Such a list is treated as if it had the same number of points as the other lists, with all the points having the same value as the one specified point.

Note

All list subsystem commands (as well as CURR:MODE LIST and VOLT:MODE LIST) are implied

 

ABORt commands.

 

 

LIST:COUN

This command sets the number of times that the list is executed before it is completed. The command accepts parameters in the range 1 through 9.9E37, but any number greater than 65534 is interpreted as INFinity. Use INF if you wish to execute a list indefinitely.

Command Syntax

[SOURce]:LIST:COUNt <NRf+>

Parameters

1 to 9.9E37 INFinity

*RST Value

1

 

Examples

LIST:COUN 3

LIST:COUN INF

Query Syntax

[SOURce]:LIST:COUNt?

Returned Parameters

<NR3>

 

Related Commands

CURR:MODE LIST:CURR LIST:DWEL LIST:STEP

 

LIST:VOLT

VOLT:MODE

LIST:CURR

This command specifies the output current points in a list. The current points are given in the command parameters, which are separated by commas. Up to 20 points may be entered and the output current values specified by the points will be generated in the same order as they were entered.

Command Syntax

[SOURce]:LIST:CURRent <NRf+> {,<NRf+>}

Parameters

See Table 3-2

Default Suffix

A

Examples

LIST:CURR 2.5,3.0,3.5 LIST:CURR MAX,2.5,MIN

Query Syntax

(None)

Related Commands

CURR:MODE LIST:CURR:POIN? LIST:DWEL

Language Dictionary 37

Image 36
Contents Agilent Part No Microfiche Part No Update April Programming GuideSafety Guidelines Contents Abor Status Reporting Synchronizing Power Module Output Changes Error MessagesIntroduction About This GuideDocumentation Summary External ReferencesVXIplug&play Power Products Instrument Drivers Downloading and Installing the DriverAccessing Online Help Supported ApplicationsGpib Capabilities Of The Power Module Introduction To ProgrammingModule Gpib Address Introduction To ScpiRST *IDN? *SRE Volt LEV Prot Curr Voltlev 8.0 Prot 8.8 CURR?Outpprotdel Traversing the Command Tree Effect of Optional HeadersOutputprotectionclear STATUSOPERATIONCONDITION? OUTPUTPROTECTIONCLEARSTATUSOPERATIONCONDITION?Voltagelevel 7PROTECTION 8CURRENTLEVEL 3MODE List Volttrig 7.5INIT*TRG Outp OFF*RCL 2OUTP onSymbol Numerical Data Formats Talking FormatsListening Formats Suffixes and Multipliers Class Unit Unit with MultiplierSystem Considerations Assign @PM3TOError Handling Agilent Basic ControllersUsing the National Instruments Gpib Interface Sending the Command Volt 5 in Basic Sending the Command Volt 5 in CReceiving Data from the Module Receiving Module Data with BasicReceiving Module Data with C Language Dictionary IntroductionDescription Of Common Commands CLSMeaning and Type DescriptionBit Configuration of Standard Event Status Enable Register ESEESE ESR?IDN? OPCOPC? OPT? PSCRCL RCLRST SAV SRESRE Bit Configuration of Status Byte Register STB?TRG Description of Subsystem Commands TST?WAI AborSubsystem Tree Diagram Calibration Subsystem CalautoCalauto 1 Calauto Once CalcurrCalcurrlev CalpassCalsave CalstatCalvolt CalvoltlevCalvoltprot CurrCurr 500 MA Currlev CurrmodeCurrprotstat Currprotstat OFFCurrtrig Currtrig 1200 MA CurrlevtrigInit Initcont 1 Initcont on ListcounListcoun Listcoun INF ListcurrLISTCURRPOIN? ListdwelLISTDWELPOIN? ListstepListvolt Listvolt 2.0,2.5,3.0 Listvolt MAX,2.5,MINLISTVOLTPOIN? MEASCURR? MEASVOLT?Outp Outpstat ON,NORELAY OutpprotOutpprotcle Outpprotdel 75E-1 OutprelOutprelpol NormOutpttlt Outpttlt 1 Outpttlt OFFOutpttltlink Outpttltsour LinkSTATOPER? STATOPEREVEN?STATOPERCOND? StatoperenabStat Oper NTR 32 Stat Oper PTR StatpresSTATQUES? STATQUESEVEN?STATQUESCOND? Stat Ques COND?SYSTERR? SYSTVERS?Trigger Subsystem TrigTrig Trig IMM TrigdelTrigdel .25 Trigdel MAX TriglinkVolt VoltlevVoltmode Voltmode List Voltmode FIXVOLTSENSSOUR? VolttrigVolttrig 1200 MV Voltlevtrig Power Module Programming Parameters Link Parameter ListPower Module Status Structure Status Register Bit ConfigurationStatus Reporting Operation Status GroupBit Signal Bit Configurations of Status Registers Meaning Status Questionable Commands QueryQuestionable Status Group Power Module Status Model Standard Event Status GroupStatus Byte Register Output QueueLocation Of Event Handles Initial Conditions At Power On ExamplesStatoperptr 5376ENAB Statquesptr 18ENABSTATOPEREVEN?QUESEVEN? Statoperptr 1024NTR Statoperenab 1024*SRESynchronizing Power Module Output Changes Trigger SubsystemModel of Fixed-Mode Trigger Operation Delaying State Idle StateInitiated State Model of List Mode Trigger Operation INITiateCONTinuous CommandTrigger Status and Event Signals Output Change StateOutpttltsour List Subsystem Listvolt 3.0,3.25,3.5,3.75 Listdwel 10,10,25,40 Automatically Repeating a ListListcurr 2,3,12,15 Triggering a ListTiming diagrams of Liststep Operation DFI Discrete Fault Indicator Subsystem Scpi Command CompletionRI Remote Inhibit Subsystem Error Messages Power Module Hardware Error MessagesSystem Error Messages Standard Event Status Register Error Bits222 -223 -241 -310 -330 -350 -400 -410 -420 -430 Scpi Confirmed Commands Scpi Approved CommandsScpi Conformance Information Scpi VersionNon-SCPI Commands Application Programs Application 1. Sequencing Multiple Modules During Power Up Figure B1-1. Block Diagram of Application #1 Variations On This ImplementationFigure B1-2. Timing Diagram of Application #1 Reset and Clear Module Enable Backplane TTL Trigger DriveEnable Output Enable Response to TriggerImplementation Details How The MPS Implements The Solution MPS Set Up Figure B2-1. Block Diagram of Application #2 When a CV-TO-CC Transition Occurs Enable TTL Trigger DriveEnable Response to TTL Trigger Start AT 15Application 3. Controlling Output Voltage Ramp Up at Turn On Figure B3-1. Simulating a Slow Voltage Ramp Generating the Desired Voltage Ramp for Application #3 Option Base Start Voltage for RampStop Voltage for Ramp SecondsApplication 4. Providing Time-Varying Voltages Figure B4-1. Voltage Waveform for Application #4Module set up Variations On This Implementation Enable OCP No Delay Before Protection OccursEnable Detection of OC Condition Enables Detection on Positive TRANSITION, I.EApplication 5. Providing Time-Varying Current Limiting Figure B5-1. Typical DUT Current vs. Time Implementation Details How The MPS Implements The Sequence GO to 12 V When Triggered Current Limit DataDwell Time Data SET to GET Current from ListApplication 6. Output Sequencing Paced by the Computer Nominal 12MPS Set Up Figure B6-1. Block Diagram of Application #6 These are the Bias Supply Limit ConditionsTo be Tested Number of Bias Supply Limit C0MBINATIONSReturn Overview Of Application Advantages/Benefits Of The MPS Solution Figure B7-1. Block Diagram of Application #7 When the Module Indicates SIC Step Completed When IT Completes the LIST. OPC Generates SROEnable SRQ Interrupt Enable Intr Identify Handler SubroutineSupplemental Information CMD$ = Output on ‘ Enable Output CMD$ = Initiate ‘ Enable Trigger to Start List‘ Conversion to Send Real Numbers Over the BUS CMD$ = Voltmode List ‘ SET to GET Voltage from ListCall Iooutputs SLOTO, CMDS, L Waiting for Trigger BIT 5 of the Operation Status RegisterCONDITION.DATA = Wend‘ Disable Auto Serial Poll If IBSTA% 0 then Goto ‘ AS Part of the Command String‘ Program N3.BAS ‘ INSTRUMENT.NAME$ = Sloto‘ General Error Handler If IBSTA% 0 then Goto Selected AS a Trigger SourceStop Application #3 Controlling Voltage Ramp UP AT Turn on Dwell = ramptimeTo terminate the iooutputa Int error Char *badstring If error != EOl enabled for both read and write Strcatvlist, vpoint This is a generalized error checking routine Index IndexIndex Index Index Agilent Sales and Support Offices United States Latin AmericaManual Updates