Model 90100
Operators Manual
The 35360A TRACKER Display supports four units of radiation. To convert between these units, use Table
Table
To Convert: | To: | Multiply By: |
|
|
|
roentgen (R) | gray (Gy) | .00879 |
|
|
|
roentgen (R) | sievert (Sv) | .00879 |
|
|
|
roentgen (R) | rad | 0.879 |
|
|
|
gray (Gy) | roentgen (R) | 113.77 |
|
|
|
gray (Gy) | sievert (Sv) | 1 |
|
|
|
gray (Gy) | rad | 100 |
|
|
|
sievert (Sv) | roentgen (R) | 113.77 |
|
|
|
sievert (Sv) | gray (Gy) | 1 |
|
|
|
sievert (Sv) | rad | 100 |
|
|
|
rad | roentgen (R) | 1.1377 |
|
|
|
rad | gray (Gy) | .01 |
|
|
|
rad | sievert (Sv) | .01 |
|
|
|
DISPLAY CALIBRATION ITEMS
Charge Gain (QG)
The 35360A TRACKER Display uses the Charge Gain to convert the number from the
Current Gain, Current Zero (IG and ADZ)
The 35360A TRACKER Display uses the Current Gain and Current Zero to convert the number from the
Ion Chamber Calibration Factor (ICCF)
The 35360A TRACKER Display uses the Ion Chamber Calibration Factor to convert charge from the 35300A TRACKER Detector ion chambers to radiation units. The factory calibrates the 35300A TRACKER Detector in units of roentgens per nanocoulomb. The accuracy of the calibration factor is ± 2.5%.
5.7 Setting the Electrometer Gains and Offsets
To set the electrometer gains, you must have a Keithley Instruments, Inc. Model 263 Calibrator/Source or equivalent. The current required for calibration procedure below must be 15 nanoamperes ± 0.07%; the charge required for calibration procedure below must be 950 nanocoulombs ± 0.5%. The Electrometer Offsets must be specified before the Electrometer Gains.