Fluke 90100 manual Changing the ION Chamber Calibration Factor Units

Page 55

Customization

Usage Examples 5

CHANGING THE ION CHAMBER CALIBRATION FACTOR UNITS

CAUTION

This example changes the current value of a factory-set calibration factor. Make a note of the value BEFORE changing it so that it may be set to this value at the end of the example.

1.Press the Page Down key until the Display Calibration Items screen appears.

2.Move the cursor to the Ion Chamber Calibration Factor for the Center chamber.

3.IMPORTANT: Make a note of the current value.

4.Press F2. The field clears.

5.Enter 1.234 and press Enter.

6.The software indicates that it is downloading. When the Item reappears, the download was successful. The number that appears is the value returned from the 35360A TRACKER Display.

7.Press F2.

8.Enter the original value and press Enter.

9.The software restores the original value in the instrument.

CHANGING THE CURRENT GAIN

CAUTION

This example changes the current value of a factory set current gain. Make a note of the value BEFORE changing it so that it may be set to this value at the end of the example.

1.Press the Page Down key until the Display Calibration Items screen appears.

2.Move the cursor to the Current Gain for the Top chamber.

3.IMPORTANT: Make a note of the current value.

4.Press F2. The field clears.

5.Enter 1e100 and press Enter.

6.The software indicates that it is downloading. Because the value, 1e100, is not a valid value for the Current Gain, the software indicates a mismatch error. Press any key to clear the error. Note that the instrument returned a value but not the one that was entered.

7.Enter the original value and press Enter.

8.The software restores the original value in the instrument.

5-9

Image 55
Contents Tracker System ModelFluke Biomedical Table of Contents Customization Maintenance and ServiceAppendix a Using Build-Up Theory of OperationIntroduction General InformationModel FeaturesManual Addenda Before You Start Safety PrecautionsSafety Terms SpecificationsIon Chamber Factory Calibration Nominal Ion Chamber CharacteristicsMechanical Specifications Recommended Operation ConditionsElectrometer / A-D Converter Measurements ModeDisplay Mode Model 35360A Tracker DisplayEnvironment High Performance Battery ChargerUnpacking and Inspection Interface Cable Detector to DisplayOptional Equipment Standard System ContentsBlank Rear Panel Familiarization OperationDetector Input Powering the SystemDisplay. Be sure to use the correct end of the cable 3 RS-232 Computer Interface RJ-45 Style Positioning the Model 35300A Detector in the BeamOperation Positioning the Model 353300A Detector in the Beam Front Panel Controls Operation InstructionsUnits Select PowerReset Test/DetectorPowering Down the Model 35360A Tracker Display Powering Up the Model 35360A Tracker DisplayUsing the TEST/DETECTOR Button Test/Detector Screen 2 Ion Chamber Calibration Factors Using the Units Select ButtonUsing the Gain Select Button Normal Mode Using the AIR Density ButtonAbsolute Measurements Using the Measure Select Button Normal ModePercent-of-Center Measurements Underranging and Overranging Conditions Battery and Bias Annunciators24. Positions of the Battery and Bias Annunciators Gain Setting ModeAborting the Fpgf Changes Saving the Fpgf ChangesBlank Theory of Operation Hardware Block Diagram of the Model 35360A Tracker Display Power SupplyVFD Display Front Panel Switches2 300-VOLT Electronic Bias Supply ProcessorElectrometer Serial Communications PortMultiplexer and A/D Converter Air Density Corrections Model 35300A Ion Chamber DetectorsBasic Operating Equations Basic Operating Quantities Blank Maintenance & Service Problem TroubleshootingNo Display Flowchart Low battery Yes Annuciator displayed? Low Battery Indication at the Display35360A Tracker Bias Error LoBias or HiBias Indication at the DisplayLeakage Flowchart Dose Rate Measurement Zero Problem Flowchart Line Fuse ReplacementIf fuses continue to blow return instrument for repair Cleaning the Instrument Interface Connector CleaningReplaceable Parts Information Recalibration & Repair Information / Technical Assistance Replaceable Parts ListCustomization Screen Formats Make a Working DiskConnection to the Computer Starting the Customization SoftwareDisplay Customization Items Screen Appearance F10 Exit F1 Help On-Line Help for using the Customization softwareF2 Edit Item Editing F6 Print Report PrintingDefinitions Ion Chamber Calibration Factor Iccf Setting the Electrometer Gains and Offsets Charge Gain QG Current Gain, Current Zero IG and ADZElectrometer Charge Gains QG Electrometer Current Gain IGUsage Examples Changing the Power Down IntervalChanging the ION Chamber Calibration Factor Units Input Errors ErrorsBuild-Up, Peg, and Retainer Kit Options Using Build-UpNominal Number of build-up Solid Water Appendix Fluke Biomedical