Fluke 90100 manual Basic Operating Quantities

Page 35

 

 

 

 

Theory of Operation

3

 

 

 

 

Basic Operating Quantities

3.6 Basic Operating Quantities

 

 

 

Table 3-3. Basic Operation Quantities

 

 

 

 

 

 

 

 

 

 

 

Quantity

Description

Units

Comments

 

 

 

 

 

 

 

 

 

 

 

 

 

Electrical current delivered to readout in

 

 

 

I

Current

nC/Min

response to radiation incident upon

 

 

 

 

 

 

detector.

 

 

 

 

 

 

Electrical charge delivered to readout in

 

 

 

Q

Charge

nC

response to radiation incident upon

 

 

 

 

 

 

detector.

 

 

 

 

 

 

Electrometer output level returned by

 

 

 

ADC

A-D Counts

Counts

A-D Converter during dose or dose rate

 

 

 

measurement.

 

 

 

 

 

 

 

 

 

 

 

 

Electrometer output level returned by

 

 

 

 

 

 

A-D Converter while operating in the

 

 

 

 

 

 

dose rate measurement mode and in

 

 

 

ADZ

A-D Zero

Counts

the absence of radiation. A-D Zero

 

 

 

levels are field programmable via

 

 

 

 

 

 

 

 

 

 

 

 

Customization Software.

 

 

 

 

 

 

Overall electrical charge gain of the

 

 

 

 

 

 

electrometer / A-D combination when

 

 

 

 

 

 

measuring in the dose mode. Charge

 

 

 

QG

Charge Gain

nC / Count

gain values are field programmable via

 

 

 

 

 

 

Customization Software.

 

 

 

 

 

 

Overall electrical current gain of the

 

 

 

 

 

 

electrometer and A-D combination

 

 

 

 

 

 

when measuring in the dose rate mode.

 

 

 

IG

Current Gain

(nC/Min) / Count

Current gain values are field

 

 

 

programmable via Customization

 

 

 

 

 

 

 

 

 

 

 

 

Software.

 

 

 

 

 

 

Specific R/nC values are supplied by

 

 

 

 

Ion Chamber

Ion Chamber

factory for each of the five ion chamber

 

 

 

 

channels. Values are field

 

 

 

ICCF

Calibration

Calibration Units

 

 

 

programmable via Customization

 

 

 

 

Factor

/ nC

 

 

 

 

Software.

 

 

 

 

 

 

 

 

 

DUCF

Display Units

Display Units

Factory default of 0.879 rad/R is

 

 

 

Conversion

/ Radiation Calibration

supplied. Value is field programmable

 

 

 

 

Factor

Units

via Customization Software.

 

 

 

 

Air Density

 

Calculated by instrument and depends

 

 

 

ADCF

Correction

None

on front panel Temperature and

 

 

 

 

Factor

 

Pressure settings.

 

 

 

FPGF

Front Panel

None

User specified via front panel gain

 

 

 

Gain Factor

setting.

 

 

 

 

 

 

 

3-7

Image 35
Contents Tracker System ModelFluke Biomedical Table of Contents Customization Maintenance and ServiceAppendix a Using Build-Up Theory of OperationIntroduction General InformationModel FeaturesManual Addenda Before You Start Safety PrecautionsSafety Terms SpecificationsIon Chamber Factory Calibration Nominal Ion Chamber CharacteristicsMechanical Specifications Recommended Operation ConditionsElectrometer / A-D Converter Measurements ModeDisplay Mode Model 35360A Tracker DisplayEnvironment High Performance Battery ChargerUnpacking and Inspection Interface Cable Detector to DisplayOptional Equipment Standard System ContentsBlank Rear Panel Familiarization OperationDetector Input Powering the SystemDisplay. Be sure to use the correct end of the cable 3 RS-232 Computer Interface RJ-45 Style Positioning the Model 35300A Detector in the BeamOperation Positioning the Model 353300A Detector in the Beam Front Panel Controls Operation InstructionsUnits Select PowerReset Test/DetectorUsing the TEST/DETECTOR Button Powering Up the Model 35360A Tracker DisplayPowering Down the Model 35360A Tracker Display Test/Detector Screen 2 Ion Chamber Calibration Factors Using the Units Select ButtonUsing the Gain Select Button Normal Mode Using the AIR Density ButtonAbsolute Measurements Using the Measure Select Button Normal ModePercent-of-Center Measurements Underranging and Overranging Conditions Battery and Bias Annunciators24. Positions of the Battery and Bias Annunciators Gain Setting ModeAborting the Fpgf Changes Saving the Fpgf ChangesBlank Theory of Operation Hardware Block Diagram of the Model 35360A Tracker Display Power SupplyVFD Display Front Panel Switches2 300-VOLT Electronic Bias Supply ProcessorMultiplexer and A/D Converter Serial Communications PortElectrometer Air Density Corrections Model 35300A Ion Chamber DetectorsBasic Operating Equations Basic Operating Quantities Blank Maintenance & Service Problem TroubleshootingNo Display Flowchart Low battery Yes Annuciator displayed? Low Battery Indication at the Display35360A Tracker Bias Error LoBias or HiBias Indication at the DisplayLeakage Flowchart Dose Rate Measurement Zero Problem Flowchart Line Fuse ReplacementIf fuses continue to blow return instrument for repair Replaceable Parts Information Interface Connector CleaningCleaning the Instrument Recalibration & Repair Information / Technical Assistance Replaceable Parts ListCustomization Screen Formats Make a Working DiskConnection to the Computer Starting the Customization SoftwareDisplay Customization Items Screen Appearance F10 Exit F1 Help On-Line Help for using the Customization softwareF2 Edit Item Editing F6 Print Report PrintingDefinitions Ion Chamber Calibration Factor Iccf Setting the Electrometer Gains and OffsetsCharge Gain QG Current Gain, Current Zero IG and ADZElectrometer Charge Gains QG Electrometer Current Gain IGUsage Examples Changing the Power Down IntervalChanging the ION Chamber Calibration Factor Units Input Errors ErrorsBuild-Up, Peg, and Retainer Kit Options Using Build-UpNominal Number of build-up Solid Water Appendix Fluke Biomedical