General Information
Specifications 1
1.6.2 Model 35360A TRACKER Display
Measurements Mode
The measurement mode is
Dose (rad): Electrometer is operated in integrating mode with capacitive feedback.
Dose Rate (rad/min): Electrometer is operated in
Display Mode
The display mode is
Absolute: All five ion chamber measurement values are displayed in absolute dose measurement units (typically rads or rad/min) as defined by the dose measurement unit's description.
Four peripheral ion chamber measurement values (top, bottom, left, and right) are displayed as a percentage of the center ion chamber measurement value.
Electrometer / A-D Converter
Nominal Full Scale Output: 3 volts.
A/D Converter Resolution: 16 bits unipolar.
Nominal Full Scale and Sensitivity in Electrical Units:
Full Scale | Sensitivity | Full Scale | Sensitivity |
1000.0 nC ± 2% | 0.015 nC/count | 1000.0 nC/min ± 2% | 0.015 (nC/min)/count |
|
|
|
|
An overrange indication will be displayed when the input current or charge have exceeded the
Long Term Stability: Expected to be less than 0.25% of reading per year.
Display Resolution: 0.003% of range
1
This relationship insures that the resolution of the measurement display values is as high as possible without exceeding the inherent resolving capability of the
Gain Temperature Stability: Less than ± 0.4% of reading over operating range of 10°C to 35°C.
Measurement Noise: Charge and Current measurement noise are less than ± 1 count at the display exclusive of source noise effects. Typical effects from a Co60 source are on the order of ± 0.2 rad/min
at 100 rad/min. Source noise effects during dose rate measurements on a linear accelerator may be excessive.
Measurement Accuracy: Fluke Biomedical factory calibration will provide a basic Charge (Current) measurement accuracy of ±1% of reading ± 0.3 nC (± 0.3 nC /min) exclusive of
Current Leakage: Less than 1 pA (typically 1 rad in 5 minutes)
Zero Offset Voltage: Factory trimmed to 0 counts at the display.