• Overall MO adjustment items
Item No. | Contents | |
112 |
| |
113 | MO electrical offset adjustment | |
114 | ||
| ||
118 |
| |
|
| |
221 | Low reflective CD TE gain adjustment | |
|
| |
223 | Low reflective CD TE offset adjustment | |
232 | ||
| ||
|
| |
236 | Low reflective CD ABCD level adjustment | |
|
| |
244 | Low reflective CD focus gain adjustment | |
|
| |
245 | Low reflective CD tracking gain adjustment | |
|
| |
121 | MO TE gain adjustment | |
|
| |
122 | MO TE offset adjustment | |
|
| |
134 | MO TWPP gain adjustment | |
|
| |
131 | MO x2 speed read TWPP offset adjustment | |
132 | ||
| ||
|
| |
136 | MO ABCD level adjustment | |
|
| |
144 | MO focus gain adjustment | |
|
| |
145 | MO tracking gain adjustment | |
|
| |
434 | MO write TWPP gain adjustment | |
|
| |
431 | MO write TWPP offset adjustment | |
432 | ||
| ||
|
| |
436 | MO write ABCD level adjustment | |
|
| |
445 | MO write tracking gain adjustment | |
|
| |
411 | MO x1 speed read TWPP offset adjustment | |
412 | ||
| ||
|
| |
448 | 32 cluster full recording | |
|
| |
141 | MO focus bias adjustment | |
|
|
13.To clear the data in overall adjustment mode, set the manual mode and change the mode number 021 (Res NV) to reset the NV. (See page 14)
14.When both CD and MO overall adjustments are OK, set the manual mode and clear the clock data.
For the microprocessor version 1.20 or later, set the mode num- ber 043 (Resume).
Note: In the mode number 043 (Resume), “Res***” display cannot be seen because of too fast processing.
If microprocessor version 1.20 or later
LCD display
043R e s u m e
rPress the X key, historical data clear
043R e s C l r
Note: In step 10, set the clock data to 99Y11M11D11H11M00S, and in step 14, do not set the clock.
Connection :
laser
power meter
Optical
digital voltmeter
MAIN board
Between TP1001
and TP1002
Check Method :
1.Select the manual mode of test mode, and set the laser power adjusting mode. (mode number 010)
2.Press the . key continuously until the optical
3.Open the cover and set the laser power meter on the objective lens of the optical
4.Press the N key, and set the laser MO read adjustment mode. (mode number 011)
5.Check that the laser power meter reading is 0.81 ± 0.08 mW.
6.Check that the voltage between TP1001 and TP1002 at this time is below 44 mV.
7.Press the N key, and set the laser CD read adjustment mode. (mode number 012)
8.Check that the laser power meter reading is 0.97 ± 0.10 mW.
9.Check that the voltage between TP1001 and TP1002 at this time is below 44 mV.
10.Press the N key, and set the laser MO write adjustment mode. (mode number 013)
11.Check that the laser power meter reading is 4.95 ± 0.50 mW.
12.Check that the voltage between TP1001 and TP1002 at this time is below 80 mV.
13.Press the x key.
14.Release the test mode.