Installation

3.4 Flow-Type Mounting

The cross-channel or guard-tube hole in the cell must always be covered by the solution and the solution level must be 1-1/2 inches above these holes. When mounting the cell in a pipe tee such as shown in Figure 3-1,have the solution enter the tee from below and exit to the side. As shown, the guard-tube hole is in line with the horizontal pipe run. However, if it is possible that the pipe line will not be full at all times, locate the hole just below the exit pipe to insure flooding of the cell under all conditions. As shown in Figure 3-1,always locate the cell on the pressure side, not the vacuum side of the pump. The flow-cell- housing, an accessory part having 3/4” male inlet and female outlet threads, can be used for an in-line measurement or in a bypass line as shown in Figure 3-1,depending upon the flow volume or pipe size. Adapter bushings are available to convert inlet and outlet fittings to 1/4” female threads. See Section 0 . The cell must be covered by the solution at all times. Therefore, make certain the lowest solution head is higher than the cell location. See that an air bubble does not prevent the cell from filling properly.

Flow-cell housing can be used “in-line” only if a maximum flow of 2 gallons per minute can be tolerated. To avoid cracking the 276127 flow-cell housing, use Teflon tape on cell threads and tighten cell only enough to prevent leakage.

To install, tighten the cell into a 1” schedule 40 pipe tee. If the flow-cell housing is used, assemble the cell and housing and install it in the process flow line or in a bypass line.

 

Preferred

Process

Cell Locations

Cooler

 

 

15"

 

(381mm)

 

Pump

 

a/n 23383

Figure 3-1 Typical Conductivity Measuring Installation

3.5 Immersion-Type Mounting for 04905 Series Cells

For use in applications where temperature does not exceed 85°C. The cell must be immersed to a level above the cross-channel or guard tube hole and must be immersed to 1-1/2 inches above this hole if an integral compensator is used. For most immersion applications, a 1/2” support pipe, preferably CPCV must be threaded into the cell bushing, using Teflon tape to seal the threads, thus permitting adequate immersion. Unless this pipe extension is used, do not immerse the top of the bushing. To insured that a representative sample is measured at all times, the solution must circulate through the channels. In quiescent solutions, provide sufficient agitation.

To install the cell, determine the length of 1/2” pipe required to give the immersion needed to keep the cell completely immersed at all times. Up to six feet of pipe can be used for the standard cell having seven feet of cable. Remove the small bushing at the top of the cell, slide it off the cable, and replace it with the 1/2- inch pipe. At the top of the pipe slide a pipe coupling and the small bushing back over the leadwire as shown in Fig. 3-1, or install a junction box to terminate the pipe.

6

4905 Series Conductivity Cells – Installation and Maintenance

12/07

Image 12
Contents Style Conductivity Cells Installation and Maintenance Manual Copyright, Notices, and Trademarks Abstract Revision NotesReferences Contents Figures Series Conductivity Cells Installation and Maintenance 12/07 Introduction OverviewIntroduction Specifications Specifications for 04905 SeriesSpecifications for 276127 Flow Chamber Installation Types of MountingFlow-Type Mounting Immersion-Type Mounting for 04905 Series CellsInsertion-Type Mounting FlowInMounting Dimensions for 04905 Series 26AWG Instrument Wiring for 4905 Cells with Integral Cable Electrical ConnectionsDirect Cell to Analyzer Installation Model 4905 Series to APT Series Analyzer/Transmitter Instrument Wiring for 4905 Cells with Quick Disconnect Cable Electrical Connections Maintenance To Clean the CellIntroduction To Check Conductivity SystemPlatinizing the Cell Electrodes Voltage and Time Limits for Platinizing CellsReplacement Parts and Accessories Extension Cables for Sensors with Quick Disconnect OptionPage Page Honeywell Process Solutions