Introduction

The cell constant is selected according to the range of the measuring instrument used and the solution measured. In general, a high-constant cell is used for solutions having low electrical resistance (high conductivity) and a low-constant cell is used for solutions having high electrical resistance (low conductivity). Automatic Temperature Compensation (ATC) during the measurement is provided by a built-in temperature sensing network located near the cross-channel or guard-tube holes.

The cells are molded from Polyethersulfone (PES) which is resistant to most corrosive chemicals over a wide range of temperatures. (A common exception is chlorinated hydrocarbons.) Sample solutions come into contact only with the above plastic and the platinum or nickel electrode surface. Any cell can be supplied with either electrode material.

2

4905 Series Conductivity Cells – Installation and Maintenance

12/07

Image 8
Contents Style Conductivity Cells Installation and Maintenance Manual Copyright, Notices, and Trademarks References AbstractRevision Notes Contents Figures Series Conductivity Cells Installation and Maintenance 12/07 Introduction OverviewIntroduction Specifications Specifications for 04905 SeriesSpecifications for 276127 Flow Chamber Installation Types of MountingFlow-Type Mounting Immersion-Type Mounting for 04905 Series CellsInsertion-Type Mounting FlowInMounting Dimensions for 04905 Series 26AWG Instrument Wiring for 4905 Cells with Integral Cable Electrical ConnectionsDirect Cell to Analyzer Installation Model 4905 Series to APT Series Analyzer/Transmitter Instrument Wiring for 4905 Cells with Quick Disconnect Cable Electrical Connections Maintenance To Clean the CellIntroduction To Check Conductivity SystemPlatinizing the Cell Electrodes Voltage and Time Limits for Platinizing CellsReplacement Parts and Accessories Extension Cables for Sensors with Quick Disconnect OptionPage Page Honeywell Process Solutions