Honeywell 70-82-25-18 manual Specifications for 04905 Series

Page 9

Specifications

2.Specifications

2.1Specifications for 04905 Series

Parameter

Description

 

 

Cell Constant

04905 Series: 0.01, 0.1, 1.0, 10 and 50 as

 

specified

Electrode Material

Nickel, Platinum or Monel as specified

 

 

Maximum Pressure Limit

1724 kPa @ 140°C (250 psig @ 284°F)

 

 

Maximum Continuous Temperature Limit

140°C (284°F)

 

For immersion applications:

 

80°C (176°F)

Materials of Construction

Cell Body: PES (polyethersulfone)

 

Support Fittings: Ryton

 

Electrodes: Nickel, Platinum or Monel as specified

 

Quick Disconnect Receptacle: Stainless Steel

 

 

Cable Options

Leadwire: PVC insulated 22 gage cable, 0.245”

 

OD, 20 and 50 feet lengths available.

 

Quick Disconnect Option

 

Mating cables must be purchased from Honeywell

 

Universal Head (Aluminum)

 

 

Weight

Approximately 1 lb (0.45 kg)

 

If using universal head:

 

3 lb (1.35 kg)

Approvals

Manufactured to comply with ASME boiler and

 

pressure vessel code Section III, Div.1, UG-101

 

CRN #0F11607.5C

 

 

Insertion

1” NPT male, Schedule 40

 

 

Flow Chamber

Inlet: 3/4" MNPT

 

Outlet: 3/4" FNPT

 

 

Insertion Depth

5” to 7” (127 to 178 mm) depending on cell

 

constant

Overall Length

Approximately 6 to 8” (152 to 203 mm)

 

If using universal head:

 

10 to 12-1/4” (254 to 311 mm)

12/07

4905 Series Conductivity Cells – Installation and Maintenance

3

Image 9
Contents Style Conductivity Cells Installation and Maintenance Manual Copyright, Notices, and Trademarks Abstract Revision NotesReferences Contents Figures Series Conductivity Cells Installation and Maintenance 12/07 Overview IntroductionIntroduction Specifications for 04905 Series SpecificationsSpecifications for 276127 Flow Chamber Types of Mounting InstallationImmersion-Type Mounting for 04905 Series Cells Flow-Type MountingFlowIn Insertion-Type MountingMounting Dimensions for 04905 Series 26AWG Electrical Connections Instrument Wiring for 4905 Cells with Integral CableDirect Cell to Analyzer Installation Model 4905 Series to APT Series Analyzer/Transmitter Instrument Wiring for 4905 Cells with Quick Disconnect Cable Electrical Connections To Clean the Cell MaintenanceIntroduction To Check Conductivity SystemVoltage and Time Limits for Platinizing Cells Platinizing the Cell ElectrodesExtension Cables for Sensors with Quick Disconnect Option Replacement Parts and AccessoriesPage Page Honeywell Process Solutions