Electrical Connections
50 FT. MAX.
WIRE. 20 FT. OR 50 FT. CABLE LENGTH.
Direct Cell to Analyzer Installation
50 FT. MAXIMUM
49054800SERIESCONDUCTIVITYCELL
Cell to Analyzer through Junction Box
NOTES:
1.FOR PURE WATER SAMPLES IN
2.FOR CELL LEADS BLACK AND WHITE, USE 16 TO 22 AWG CABLE SHIELDED TWISTED PAIR, WITH 30pF MAX. CAPACITANCE BETWEEN CONDUCTORS, CONNECT SHIELD TO TERMINAL “10”
3.FOR COMPENSATOR LEADS RED AND GREEN, USE 16 TO 22 AWG, TWO CONDUCTOR CABLE
4.CELL TO ANALYZER CABLES ARE CONSIDERED LOW LEVEL. RUN SEPARATE FROM HIGH LEVEL WIRING.
Figure 4-2 Installation Diagram, 4905 Cells, with 20′ leads directly connected to
UDA2182 Analyzer or connected to Junction Box