Honeywell 70-82-25-18 manual Specifications for 276127 Flow Chamber

Page 10

Specifications

2.2 Specifications for 276127 Flow Chamber

Parameter

Description

 

 

Maximum Flow

2 gpm @ 40psig and atmospheric discharge

 

 

Maximum Pressure

200 psig @ 25°C

 

 

Maximum Temperature

140°C (284°F) at atmospheric pressure

 

 

Dimensions

1-1/2” (3.8 cm) octagon x 8-3/4” (22.2 cm) long.

 

Sample Inlet: 3/4" MNPT

 

Sample Outlet: 3/4" FNPT

 

Cell Inlet: 1” MNPT

Materials of Construction

Polyethersulfone (PES)

 

 

4

4905 Series Conductivity Cells – Installation and Maintenance

12/07

Image 10
Contents Style Conductivity Cells Installation and Maintenance Manual Copyright, Notices, and Trademarks Revision Notes AbstractReferences Contents Figures Series Conductivity Cells Installation and Maintenance 12/07 Introduction OverviewIntroduction Specifications Specifications for 04905 SeriesSpecifications for 276127 Flow Chamber Installation Types of MountingFlow-Type Mounting Immersion-Type Mounting for 04905 Series CellsInsertion-Type Mounting FlowInMounting Dimensions for 04905 Series 26AWG Instrument Wiring for 4905 Cells with Integral Cable Electrical ConnectionsDirect Cell to Analyzer Installation Model 4905 Series to APT Series Analyzer/Transmitter Instrument Wiring for 4905 Cells with Quick Disconnect Cable Electrical Connections Introduction MaintenanceTo Clean the Cell To Check Conductivity SystemPlatinizing the Cell Electrodes Voltage and Time Limits for Platinizing CellsReplacement Parts and Accessories Extension Cables for Sensors with Quick Disconnect OptionPage Page Honeywell Process Solutions