Electrical Connections

4. Electrical Connections

4.1 Overview

The terminal board connections for the various Honeywell measuring instruments are given in the appropriate Figures in this section.

To avoid the possibility of AC pickup in the cell leads, separate them from all AC line-voltage wiring or run them in a separate grounded conduit.

ATTENTION

Do not use shielded cable except where shown in the following figures.

4.2 Instrument Wiring for 4905 Cells with Integral Cable

4.2.1 Model 4905 Series with Integral Cable to UDA2182 Analyzer

NOTES:

1.FOR PURE WATER SAMPLES IN NON -CONDUCTIVE (PLASTIC, GLASS, ETC) PIPING, GROUND THE BLACK CELL ELECTRODE LEAD NEAR THE CELL. ALTERNATELY, CONNECT TO THE UDA GROUND SCREW AS SHOWN DOTTED. DO NOT GROUND 10, 25, OR 50 CONSTANT CELLS

2.FOR CELL LEADS A AND C, USE 16 TO 22 AWG CABLE SHIELDED TWISTED PAIR,

WITH 30pF MAX. CAPACITANCE BETWEEN CONDUCTORS, CONNECT SHIELD TO TERMINAL “10”

3.FOR COMPENSATOR LEADS B AND D, USE 16 TO 22 AWG, TWO CONDUCTOR C ABLE

4. CELL TO ANALYZER CABLES ARE CONSIDERED LOW LEVEL. RUN SEPARATE F ROM HIGH LEVEL WIRING.

Figure 4-1 Installation Diagram, 4905 Cells, with Junction Box head connected to UDA2182

Analyzer

10

4905 Series Conductivity Cells – Installation and Maintenance

12/07

Image 16
Contents Style Conductivity Cells Installation and Maintenance Manual Copyright, Notices, and Trademarks Revision Notes AbstractReferences Contents Figures Series Conductivity Cells Installation and Maintenance 12/07 Introduction OverviewIntroduction Specifications Specifications for 04905 SeriesSpecifications for 276127 Flow Chamber Installation Types of MountingFlow-Type Mounting Immersion-Type Mounting for 04905 Series CellsInsertion-Type Mounting FlowInMounting Dimensions for 04905 Series 26AWG Instrument Wiring for 4905 Cells with Integral Cable Electrical ConnectionsDirect Cell to Analyzer Installation Model 4905 Series to APT Series Analyzer/Transmitter Instrument Wiring for 4905 Cells with Quick Disconnect Cable Electrical Connections Maintenance To Clean the CellIntroduction To Check Conductivity SystemPlatinizing the Cell Electrodes Voltage and Time Limits for Platinizing CellsReplacement Parts and Accessories Extension Cables for Sensors with Quick Disconnect OptionPage Page Honeywell Process Solutions