Newcon Optik LRB 6000CI operation manual Compass calibration, Setting interface format

Page 22

Compass calibration

If the System was exposed to a strong magnetic field or was not in use for a long time, the compass accuracy may diminish.

To return the compass into working condition:

Select CALc from CAL menu

Keeping the unit horizontally, perform one full rotation around vertical axis within half a minute and then one full rotation around vertical axis with device upside down during half a minute.

If one rotation does not restore calibration – perform two rotations.

The normal compass functioning will be restored.

Setting interface format

In this mode data communication standard is set. The unit can output measurement data via RS-232 port in either PC format (description of the proprietary format can be obtained from Newcon upon a request) or in the format accepted by PLGF/DALGR.

To select required data format:

Select PLGr or PC from SEt menu

Pressing Action button will activate PC or PLGR format, the display name of format that can be activated will be shown.

18

Image 22
Contents LRB 6000CI Page Important Information Read prior to activationPage Table of Contents Brief Description Key Features of the LRB 6000CI Laser Rangefinder Binoculars Device Appearance Front viewRear view Display Delivery SET Range Finder SpecificationsOptics Or Inclinometer PowerCompass EnvironmentalInstalling the battery Operation InstructionsUsing the Device Factors affecting measuring distance Target selection logic Individual Measurement and Scanning regimesOperation and service modes Page Accuracy correction Std Sd, d DCESetting interface format Compass calibrationGating mode Data Recall mode Computer interconnectivity Short mode switching scheme GPS interconnectivityBest Measuring Technique Page Storage and Maintenance Troubleshooting Black dots are visible on LCDWarranty What This Warranty Does Not CoverUS customers Customer SupportAll other countries Quality Certificate Page Page Newcon Optik