Newcon Optik LRB 3000PRO operation manual Accuracy correction, Sd, d

Page 19

Accuracy correction

If the System produces some measurement errors that lie beyond unit specification, it may be an indication that unit requires calibration similar to one conducted at production level. To do accuracy correction, please select Corr from CAL menu.

Perform correction according to required measurement parameters by pressing Action button, see table below:

Active mode of

Correction

Condition

measurement (Std)

parameter

 

 

 

Install unit at 1 meter1

S, Sd, d

Distance

from flat target2

C

Azimuth

Put unit in a level plane

and aim it to the North

 

 

direction3

 

 

Install unit at 1 meter1

dC

Distance and

from flat target2

controlling its level

 

azimuth

position and direction to

 

 

 

 

the North

1 – Accuracy of correction will depend on accuracy of installation against target. If distance is 0.9 meters, then unit will have 0.1 m offset during all future measurements.

15

Image 19 Contents
LRB 3000PRO Page Read prior to activation Important InformationPage Table of Contents Principle of operation Brief DescriptionKey Features Device Appearance Page Display Standard delivery set SpecificationsOptional accessories Delivery SETCompass PowerRange Finder Speed measurementPreparing device for operation Operation InstructionsEnvironmental MechanicsMeasurement procedure Target selection logic Factors affecting measuring distanceIndividual Measurement and Scanning regimes Operation and service modes Page Sd, d Accuracy correctionCompass calibration Data Recall mode Best Measuring Technique Storage and Maintenance Compass readings are equal to zero at any direction TroubleshootingWhat This Warranty Does Not Cover WarrantyCustomer Support Serial number Quality CertificateNewcon Optik