Invacare manual Pulse Button MA80, Alternating Pressure Button MA85

Page 15

SECTION 2—OPERATION

Pulse Button (MA80)

1.Press the Pulse button () to Turn on Pulse mode and select one of four Pulse Times - 1, 3, 5 or 10 minutes.

NOTE: The pulse times are 1, 3, 5 or 10 minutes. The pulse times indicate the time between each low pulse. For example, if 5 minutes is selected, all the air cushions in the mattress will inflate for 5 minutes (the selected time) to the selected comfort setting and then drop 20% pressure for 20 seconds.

Alternating Pressure Button (MA85)

1.Press the Alternating Pressure button () to select one of four alternating pressure modes.

NOTE: The alternating pressure times indicate the frequency of deflation or inflation of half of the air cushions (even or odd numbered). For example, 10 minutes is selected. In this example, the even numbered air cushions in the mattress will deflate, while the odd numbered cushions remain at constant pressure. When 10 minutes has elapsed, the even numbered air cushions will inflate while the odd numbered cushions remain at constant pressure. When 10 minutes has elapsed again, the odd numbered air cushions deflate while the even numbered cushions remain at constant pressure. The air cushions continue to inflate or deflate like this as long as an alternating pressure time has been selected. At all times, half of the air cushions remain at a constant pressure when an alternating pressure time has been selected.

Adjusting the Minimum Pressure Setting (A/P Low)

The minimum pressure setting during Alternating Pressure is set to 0 (2 mmHg) at the factory. This means that the mattress will deflate as much as possible during the deflation time when using 10, 15 or 20 minute Alternating Pressure times. The setting can be changed so the mattress will only deflate to 50% of the selected pressure setting.

1.Ensure the Standby LED is lit.

2.Press and hold the Alternating Pressure () and Fowler ( ) buttons.

3.Read the LCD Display. The LCD Display will show:

0 if the current setting is 0 (2 mmHg).

5 if the current setting is 50% of the selected pressure.

NOTE: The On and Standby LEDs will be lit.

4.Press the Soft () or Firm ( ) buttons to toggle between the two settings.

5.Press the Power button to save the setting.

NOTE: The control unit will exit this adjustment mode if no button is pressed for 30 seconds.

Part No 1148138

15

MA80 Series

Image 15
Contents Dealer This manual Must be given to the user of the product MA80 SeriesSpecial Notes Special NotesTable of Contents Register Your Product Register Online at warranty.invacare.comMechanical Parameters Typical Product ParametersElectrical Parameters Performance ParametersEnvironmental Parameters Label Location Entrapment May Occur InstallationGeneral Guidelines ContraindicationsAnesthesia Equipment Fire HazardElectrical SmokingOperation Installing the Mattress Replacement SystemInstalling the Mattress Replacement System Installing the Side Rails Installing the Control UnitConnecting the Hose Connecting the HoseConnecting the Power Cord Connecting the Power CordPower Button Using the Front PanelAlternating Pressure Button MA85 Pulse Button MA80Static Button Firm/Soft ButtonsFowler Button Low Pressure LED Lock/Alarm Silence ButtonPower Fail LED Max Inflate ButtonPlacing the Patient on the Mattress Powering Up the SystemPreparing for CPR Procedure Transferring Patient From/To a GurneyTransferring Patient From/To a Wheelchair Preparing for CPR Procedure Cleaning the System Maintenance and TroubleshootingStoring the System Changing the Filter TroubleshootingUSA Limited Warranty