SECTION 2—OPERATION

Static Button

1.Press the Static button () to select the static mode and maintain the mattress at constant pressure.

NOTE: The Static LED is illuminated GREEN.

Firm/Soft Buttons

1.Select comfort pressure settings by pressing Firm ( ) or Soft () buttons.

Soft Button - Pressing this button reduces the pressure in the mattress.

Firm Button - Pressing this button increases the pressure in the mattress.

NOTE: The patient comfort pressure ranges from Soft (level 0 = 6±4 mmHg) to Firm (level 9 = 32±6 mmHg). The Comfort Control LED displays the patient comfort pressure levels from 0 to 9 and provides a guide to the caregiver to set approximate comfort pressure level depending on the patient weight. If the patient’s weight to height ratio is above average, increase the pressure setting by approximately 20%.

Fowler Button

Manual Mode

1. Press the Fowler button ( ) to activate the patient fowler mode.

NOTE: The Fowler LED illuminates when in this mode. When this mode is activated, the control unit increases the pressure in the mattress to prevent the patient from bottoming out.

Automatic Wireless Mode

When the bed articulates to 45°, the transmitter in the head of the mattress signals the control unit to increase the pressure in the mattress by 80% to prevent the patient from bottoming out.

Fowler Transmitter Setup

1.Make sure the bed is in the flat position.

2.Make sure fowler transmitter is in the mattress cover pocket in the orientation indicated on the transmitter.

NOTE: The mattress cover pocket is on the inside of the lower mattress cover on the left side of the head end of the mattress.

3. Press and hold the Fowler ( ) and Max Inflate () buttons.

NOTE: The Fowler LED is lit and L displays in the LCD Display.

4. Raise the bed to 45°.

NOTE: The control unit will beep and will return to Standby.

MA80 Series

16

Part No 1148138

Image 16
Contents MA80 Series Dealer This manual Must be given to the user of the productSpecial Notes Special NotesTable of Contents Register Online at warranty.invacare.com Register Your ProductTypical Product Parameters Electrical ParametersPerformance Parameters Mechanical ParametersEnvironmental Parameters Label Location Installation General GuidelinesContraindications Entrapment May OccurFire Hazard ElectricalSmoking Anesthesia EquipmentInstalling the Mattress Replacement System OperationInstalling the Side Rails Installing the Control Unit Installing the Mattress Replacement SystemConnecting the Hose Connecting the Hose Connecting the Power Cord Connecting the Power CordUsing the Front Panel Power ButtonPulse Button MA80 Alternating Pressure Button MA85Firm/Soft Buttons Static ButtonFowler Button Lock/Alarm Silence Button Power Fail LEDMax Inflate Button Low Pressure LEDPowering Up the System Placing the Patient on the MattressTransferring Patient From/To a Gurney Preparing for CPR ProcedureTransferring Patient From/To a Wheelchair Preparing for CPR Procedure Maintenance and Troubleshooting Cleaning the SystemStoring the System Troubleshooting Changing the FilterLimited Warranty USA