Invacare MA80, MA85 Typical Product Parameters, Electrical Parameters, Performance Parameters

Page 5

TYPICAL PRODUCT PARAMETERS

TYPICAL PRODUCT PARAMETERS

Electrical Parameters

 

MA80/MA85

 

 

INPUT VOLTAGE AC:

90 V

 

 

INPUT FREQUENCY:

60 Hz

 

 

CURRENT:

2 A

 

 

MAXIMUM POWER CONSUMPTION:

150 ± 10 W

 

 

CIRCUIT PROTECTION:

Dual fused,

 

250 V,

 

5 A fast blow fuses

 

 

MODE OF OPERATION:

Continuous

 

 

Performance Parameters

 

MA80

 

MA85

 

 

 

 

WEIGHT CAPACITY

 

 

 

STANDARD MATTRESS:

350 lbs

BARIATRIC MATTRESS:

1000 lbs

PRESSURE ZONE:

 

5

MAXIMUM FLOW:

1275 ± 100 LPM

MAXIMUM PRESSURE:

35 ± 5 mmHg

MAXIMUM FLOW TIMER:

15 minutes

SUPPORT SURFACE INFLATION TIME:

25 - 60 seconds

 

 

 

 

PATIENT COMFORT CONTROL PRESSURES:

 

 

 

SOFT PRESSURE:

8 ± 4 mmHg

FIRM PRESSURE:

32 ± 4 mmHg

 

 

 

 

CYCLE TIME:

1, 3, 5, 10 minutes

 

5, 10, 15, 20 minutes

 

 

 

PATIENT CONTACT:

Control unit and mattress have

 

Latex-Free components

 

 

 

 

Mechanical Parameters

 

MA80/MA85

 

 

CONTROL UNIT

 

DIMENSIONS (LXWXH):

15” x 6.5” x 11”

WEIGHT:

15 lbs

POWER CORD:

10 - 14 Feet Long, Hospital Grade

CONNECTION:

Triple Flow - Two ½” and One ¼” Single

 

Magnetic Quick Coupling

PACKAGING:

1 piece/box

 

 

AIR FILTER:

Charcoal Air Filter with Fire Retardant

 

 

Part No 1148138

5

MA80 Series

Image 5
Contents Dealer This manual Must be given to the user of the product MA80 Series Special Notes Special NotesTable of Contents Register Your Product Register Online at warranty.invacare.comElectrical Parameters Typical Product ParametersPerformance Parameters Mechanical ParametersEnvironmental Parameters Label Location General Guidelines InstallationContraindications Entrapment May OccurElectrical Fire HazardSmoking Anesthesia EquipmentOperation Installing the Mattress Replacement SystemInstalling the Mattress Replacement System Installing the Side Rails Installing the Control UnitConnecting the Hose Connecting the HoseConnecting the Power Cord Connecting the Power CordPower Button Using the Front PanelAlternating Pressure Button MA85 Pulse Button MA80Fowler Button Static ButtonFirm/Soft Buttons Power Fail LED Lock/Alarm Silence ButtonMax Inflate Button Low Pressure LEDPlacing the Patient on the Mattress Powering Up the SystemTransferring Patient From/To a Wheelchair Preparing for CPR ProcedureTransferring Patient From/To a Gurney Preparing for CPR Procedure Cleaning the System Maintenance and TroubleshootingStoring the System Changing the Filter TroubleshootingUSA Limited Warranty