SECTION
Powering Up the System
NOTE: For this procedure, refer to FIGURE 2.5 on page 14.
1. Turn on the power to the system by pressing the Power button on the control unit. NOTE: Once the button is released, a GREEN LED illuminates when the unit is on.
Placing the Patient on the Mattress
1. Press the Max Inflate button to turn on the control unit to maximum flow. NOTE: In this mode, the Max Inflate LED lights up.
2.After the mattress is fully inflated, place the patient on the bed.
3.Ensure that the patient’s feet are towards the foot end of the mattress (the end with the hose).
4.Center the patient on the bed from
NOTE: Special positioning may be required with contracted patients to provide comfortable positions.
5.After placing the patient, make certain no objects will fall under the patient, such as feeding tubes, IV's etc.
6.Wait five minutes for the mattress pressure to stabilize.
7.Set the comfort pressure to the desired comfort level.
8.Verify that the patient has not bottomed out by performing the following steps:
A.Ensure that the patient is lying flat on his/her back in the middle of the mattress.
B.Place four fingers between the air cushions directly underneath the sacral region of the patient’s body.
C.Ensure that there is 3 to
D.Adjust the comfort setting, if needed.
E.Wait five minutes for the mattress pressure to stabilize.
F.Repeat STEPS
9.If the patient feels that the bed is too soft/hard, press the Soft or Firm button to adjust the comfort settings.
10.Use a regular pillow to help support and stabilize the patient's head.
MA80 Series | 18 | Part No 1148138 |