Invacare MA85, MA80 manual Powering Up the System, Placing the Patient on the Mattress

Page 18

SECTION 2—OPERATION

Powering Up the System

NOTE: For this procedure, refer to FIGURE 2.5 on page 14.

1. Turn on the power to the system by pressing the Power button on the control unit. NOTE: Once the button is released, a GREEN LED illuminates when the unit is on.

Placing the Patient on the Mattress

1. Press the Max Inflate button to turn on the control unit to maximum flow. NOTE: In this mode, the Max Inflate LED lights up.

2.After the mattress is fully inflated, place the patient on the bed.

3.Ensure that the patient’s feet are towards the foot end of the mattress (the end with the hose).

4.Center the patient on the bed from side-to-side and head-to-foot.

NOTE: Special positioning may be required with contracted patients to provide comfortable positions.

5.After placing the patient, make certain no objects will fall under the patient, such as feeding tubes, IV's etc.

6.Wait five minutes for the mattress pressure to stabilize.

7.Set the comfort pressure to the desired comfort level.

8.Verify that the patient has not bottomed out by performing the following steps:

A.Ensure that the patient is lying flat on his/her back in the middle of the mattress.

B.Place four fingers between the air cushions directly underneath the sacral region of the patient’s body.

C.Ensure that there is 3 to 4-finger width clearance between the bottom of the patient and the bed frame.

D.Adjust the comfort setting, if needed.

E.Wait five minutes for the mattress pressure to stabilize.

F.Repeat STEPS A-E until patient has not bottomed out and patient comfort is achieved.

9.If the patient feels that the bed is too soft/hard, press the Soft or Firm button to adjust the comfort settings.

10.Use a regular pillow to help support and stabilize the patient's head.

MA80 Series

18

Part No 1148138

Image 18
Contents MA80 Series Dealer This manual Must be given to the user of the productSpecial Notes Special NotesTable of Contents Register Online at warranty.invacare.com Register Your ProductPerformance Parameters Typical Product ParametersElectrical Parameters Mechanical ParametersEnvironmental Parameters Label Location Contraindications InstallationGeneral Guidelines Entrapment May OccurSmoking Fire HazardElectrical Anesthesia EquipmentInstalling the Mattress Replacement System OperationInstalling the Side Rails Installing the Control Unit Installing the Mattress Replacement SystemConnecting the Hose Connecting the HoseConnecting the Power Cord Connecting the Power CordUsing the Front Panel Power Button Pulse Button MA80 Alternating Pressure Button MA85Static Button Firm/Soft ButtonsFowler Button Max Inflate Button Lock/Alarm Silence ButtonPower Fail LED Low Pressure LEDPowering Up the System Placing the Patient on the MattressPreparing for CPR Procedure Transferring Patient From/To a GurneyTransferring Patient From/To a Wheelchair Preparing for CPR Procedure Maintenance and Troubleshooting Cleaning the SystemStoring the System Troubleshooting Changing the FilterLimited Warranty USA