Invacare MA65, MA60 manual Alternating Pressure Button, Static Button

Page 15

SECTION 2—OPERATION

Alternating Pressure Button

1.Press the Alternating Pressure button () to select between one of four alternating pressure times - 5, 10, 15 and 20 minutes.

NOTE: The alternating pressure times indicate the frequency of deflation or inflation of half of the air cushions (even or odd numbered). For example, 10 minutes is selected. In this example, the even numbered air cushions in the mattress will deflate, while the odd numbered cushions remain at constant pressure. When 10 minutes has elapsed, the even numbered air cushions will inflate while the odd numbered cushions remain at constant pressure. When 10 minutes has elapsed again, the odd numbered air cushions deflate while the even numbered cushions remain at constant pressure. The air cushions continue to inflate or deflate like this as long as an alternating pressure time has been selected. At all times, half of the air cushions remain at a constant pressure when an alternating pressure time has been selected.

Adjusting the Minimum Pressure Setting (A/P Low)

The minimum pressure setting during Alternating Pressure is set to 0 (2 mmHg) at the factory. This means that the mattress will deflate as much as possible during the deflation time when using 10, 15 or 20 minute Alternating Pressure times. The setting can be changed so the mattress will only deflate to 50% of the selected pressure setting.

1.Ensure the Standby LED is lit.

2.Press and hold the Alternating Pressure () and Fowler ( ) buttons.

3.Read the LCD Display. The LCD Display will show:

0 if the current setting is 0 (2 mmHg).

5 if the current setting is 50% of the selected pressure.

NOTE: The On and Standby LEDs will be lit.

4.Press the Soft () or Firm ( ) buttons to toggle between the two settings.

5.Press the Power button to save the setting.

NOTE: The control unit will exit this adjustment mode if no button is pressed for 30 seconds.

Static Button

1.Press the Static button () to enter Static mode and maintain all air cushions in the mattress at a constant pressure.

NOTE: The Static LED is illuminated GREEN.

Part No 1148137

15

MA60 Series

Image 15
Contents Dealer This manual Must be given to the user of the product MA60 SeriesSpecial Notes Special NotesTable of Contents Register Your Product Register Online at warranty.invacare.comMechanical Parameters Typical Product ParametersElectrical Parameters Performance ParametersEnvironmental Parameters Label Location Entrapment May Occur Safety InstructionsGeneral Guidelines ContraindicationsAnesthesia Equipment Fire HazardElectrical SmokingOperation Installing the Powered MattressInstalling the Powered Mattress Installing the Side Rails Installing the Control UnitConnecting the Hose Connecting the HoseConnecting the Power Cord Power Button Using the Front PanelStatic Button Alternating Pressure ButtonFowler Button Firm/Soft ButtonsMax Inflate/Low Air Loss Button Lock/Alarm Silence ButtonLow Pressure LED Powering Up the SystemPower Fail LED Placing the Patient on the MattressTransferring Patient From/To a Gurney Preparing for CPR Procedure About Power Outage and TransportationTransferring Patient From/To a Wheelchair Cleaning the System Maintenance and TroubleshootingStoring the System Problem Cause Solution TroubleshootingUSA Limited Warranty