SECTION 2—OPERATION

Max Inflate/Low Air Loss Button

MA60 System

1.Press the Max Inflate button () to select the max inflate mode.

Max Inflate Mode - In this mode, the mattress inflates rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to the previous setting.

NOTE: It takes 5-10 minutes for the mattress to inflate fully (inflation time depends on size of mattress).

NOTE: Max Inflate mode can be manually disengaged by pressing the Max Inflate button. This will deactivate the Max Inflate LED.

NOTE: It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or patient cleaning.

MA65 System

NOTE: The Low Air Loss mode is only on the MA65 system.

1.Press the Max Inflate/Low Air Loss button () to select the max inflate mode or the low air loss mode.

Max Inflate Mode - In this mode, the mattress inflates rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to the previous setting.

NOTE: It takes 5-10 minutes for the mattress to inflate fully (inflation time depends on size of mattress).

NOTE: Max Inflate mode can be manually disengaged by pressing the Max Inflate button. This will deactivate the Max Inflate LED.

NOTE: It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or patient cleaning.

Low Air Loss Mode - In this mode, the mattress goes into on-demand low air loss relief mode.

Lock/Alarm Silence Button

1.Press the Lock/Alarm Silence button () to select one of the following modes:

Lock Mode - Holding this button locks out all control unit functions, including the Power button, to prevent any tampering with the settings. Press the button for approximately 3-5 seconds for the Lock LED to activate.

NOTE: The Lock LED illuminates when in this mode.

Part No 1148137

17

MA60 Series

Image 17
Contents Dealer This manual Must be given to the user of the product MA60 SeriesSpecial Notes Special NotesTable of Contents Register Your Product Register Online at warranty.invacare.comElectrical Parameters Typical Product ParametersPerformance Parameters Mechanical ParametersEnvironmental Parameters Label Location General Guidelines Safety InstructionsContraindications Entrapment May OccurElectrical Fire HazardSmoking Anesthesia EquipmentOperation Installing the Powered MattressInstalling the Powered Mattress Installing the Side Rails Installing the Control UnitConnecting the Hose Connecting the HoseConnecting the Power Cord Power Button Using the Front PanelStatic Button Alternating Pressure ButtonFowler Button Firm/Soft ButtonsMax Inflate/Low Air Loss Button Lock/Alarm Silence ButtonPower Fail LED Powering Up the SystemPlacing the Patient on the Mattress Low Pressure LEDTransferring Patient From/To a Gurney Transferring Patient From/To a Wheelchair Preparing for CPR ProcedureAbout Power Outage and Transportation Cleaning the System Maintenance and TroubleshootingStoring the System Problem Cause Solution TroubleshootingUSA Limited Warranty