Invacare MA65, MA60 Typical Product Parameters, Electrical Parameters, Performance Parameters

Page 5

TYPICAL PRODUCT PARAMETERS

TYPICAL PRODUCT PARAMETERS

Electrical Parameters

 

MA60/MA65

 

 

INPUT VOLTAGE AC:

90 V

 

 

INPUT FREQUENCY:

60 Hz

 

 

CURRENT:

1 A

 

 

MAXIMUM POWER CONSUMPTION:

30 ± 10 W

 

 

CIRCUIT PROTECTION:

Dual fused,

 

250 V,

 

1 A fast blow fuses

 

 

MODE OF OPERATION:

Continuous

 

 

Performance Parameters

 

MA60/MA65

 

 

WEIGHT CAPACITY

 

STANDARD MATTRESS:

350 lbs

BARIATRIC MATTRESS:

1000 lbs

PRESSURE ZONE:

2

MAXIMUM FLOW:

50 ± 15 LPM

MAXIMUM FLOW PRESSURE:

35 ± 5 mmHg

MAXIMUM FLOW TIMER:

30 minutes

SUPPORT SURFACE INFLATION TIME:

5 -10 minutes

 

 

PATIENT COMFORT CONTROL PRESSURES

 

SOFT PRESSURE:

8 ± 4 mmHg

FIRM PRESSURE:

32 ± 4 mmHg

CYCLE TIME:

5, 10, 15, 20 minutes

 

 

PATIENT CONTACT:

Control unit and mattress have Latex-Free

 

components

 

 

Mechanical Parameters

 

MA60

MA65

 

 

 

CONTROL UNIT

 

 

 

 

 

DIMENSIONS (L X W X H):

15” x 6.5” x 11”

 

 

WEIGHT:

15 lbs

 

 

POWER CORD:

10 - 14 Feet Long, Hospital Grade

 

 

 

CONNECTION:

Two ¼” Flow Couplings

Three ¼” Flow Couplings

 

 

 

PACKAGING:

1 piece/box

 

 

AIR FILTER:

None

 

 

 

 

 

 

Part No 1148137

5

MA60 Series

Image 5
Contents Dealer This manual Must be given to the user of the product MA60 SeriesSpecial Notes Special NotesTable of Contents Register Your Product Register Online at warranty.invacare.comElectrical Parameters Typical Product ParametersPerformance Parameters Mechanical ParametersEnvironmental Parameters Label Location General Guidelines Safety InstructionsContraindications Entrapment May OccurElectrical Fire HazardSmoking Anesthesia EquipmentOperation Installing the Powered MattressInstalling the Powered Mattress Installing the Side Rails Installing the Control UnitConnecting the Hose Connecting the HoseConnecting the Power Cord Power Button Using the Front PanelStatic Button Alternating Pressure ButtonFowler Button Firm/Soft ButtonsMax Inflate/Low Air Loss Button Lock/Alarm Silence ButtonPower Fail LED Powering Up the SystemPlacing the Patient on the Mattress Low Pressure LEDTransferring Patient From/To a Gurney Transferring Patient From/To a Wheelchair Preparing for CPR ProcedureAbout Power Outage and Transportation Cleaning the System Maintenance and TroubleshootingStoring the System Problem Cause Solution TroubleshootingUSA Limited Warranty
Related manuals