Invacare MA60, MA65 manual Firm/Soft Buttons, Fowler Button

Page 16

SECTION 2—OPERATION

Firm/Soft Buttons

1.Select comfort pressure settings by pressing Firm ( ) or Soft () button.

Soft Button - Pressing this button reduces the pressure setting in the mattress.

Firm Button - Pressing this button increase the pressure setting in the mattress.

NOTE: The patient comfort pressure ranges from Soft (level 0 = 6±4 mmHg) to Firm (level 9 = 32±6 mmHg). The Comfort Control LED displays the patient comfort pressure levels from 0 to 9 and provides a guide to the caregiver to set approximate comfort pressure level depending on the patient weight. If the patient’s weight to height ratio is above average, increase the pressure setting by approximately 20%.

Fowler Button

Manual Mode

1. Press the Fowler button ( ) to activate the patient fowler mode.

NOTE: The Fowler LED illuminates when in this mode. When this mode is activated, the control unit increases the pressure in the mattress to prevent the patient from bottoming out.

Automatic Wireless Mode

When the bed articulates to 45°, the transmitter in the head of the mattress signals the control unit to increase the pressure in the mattress by 80% to prevent the patient from bottoming out.

Fowler Transmitter Setup

1.Make sure the bed is in the flat position.

2.Make sure fowler transmitter is in the mattress cover pocket in the orientation indicated on the transmitter.

NOTE: The mattress cover pocket is on the inside of the lower mattress cover on the left side of the head end of the mattress.

3.Perform one of the following:

MA60 - Press and hold the Max Inflate and Fowler buttons.

MA65 - Press and hold the Mode and Fowler buttons.

NOTE: The Fowler LED is lit and L displays in the LCD Display.

4. Raise the bed to 45°.

NOTE: The control unit will beep and will return to Standby.

5.Perform one of the following:

Cancel - Press the Fowler button to exit Fowler Transmitter Setup without linking to a transmitter.

Save - Wait 60 seconds without pressing a key.

MA60 Series

16

Part No 1148137

Image 16
Contents MA60 Series Dealer This manual Must be given to the user of the productSpecial Notes Special NotesTable of Contents Register Online at warranty.invacare.com Register Your ProductTypical Product Parameters Electrical ParametersPerformance Parameters Mechanical ParametersEnvironmental Parameters Label Location Safety Instructions General GuidelinesContraindications Entrapment May OccurFire Hazard ElectricalSmoking Anesthesia EquipmentInstalling the Powered Mattress OperationInstalling the Side Rails Installing the Control Unit Installing the Powered MattressConnecting the Hose Connecting the HoseConnecting the Power Cord Using the Front Panel Power ButtonAlternating Pressure Button Static ButtonFirm/Soft Buttons Fowler ButtonLock/Alarm Silence Button Max Inflate/Low Air Loss ButtonPowering Up the System Power Fail LEDPlacing the Patient on the Mattress Low Pressure LEDTransferring Patient From/To a Gurney About Power Outage and Transportation Preparing for CPR ProcedureTransferring Patient From/To a Wheelchair Maintenance and Troubleshooting Cleaning the SystemStoring the System Troubleshooting Problem Cause SolutionLimited Warranty USA