Dakota Digital MX-5 operation manual Probe Zero, Performing a Probe-Zero

Page 17

MX-5 Ultrasonic Thickness Gauge

Probe Zero

Setting the Zero Point of the MX-5is important for the same reason that setting the zero on a mechanical micrometer is important. If the gauge is not "zeroed" correctly, all of the measurements the gauge makes will be in error by some fixed number. When the MX-5is "zeroed", this fixed error value is measured and automatically corrected for in all subsequent measurements. The MX-5may be "zeroed" by performing the following procedure:

Performing a Probe-Zero

1)Make sure the M X-5is on.

2)Plug the transducer into the MX-5. Make sure that the connectors are fully engaged. Check that the wearface of the transducer is clean and free of any debris.

3)The metal probe-disc is on the top end of the MX-5. Apply a single droplet of ultrasonic couplant to the face of this disc.

4)Press the transducer against the probe-disc, making sure that the transducer sits flat against the surface. The display should show some thickness value, and the Stability Indicator should have nearly all its bars illuminated.

5)While the transducer is firmly coupled to the probe-disc, press the PRB-0key on the keypad. The MX-5will display "Prb0" while it is calculating its zero point

6)Remove the transducer from the probe-disc.

13

Image 17
Contents P-114-0002 Page Disclaimer ContentsBlank Introduction Blank Keypad OperationDakota Ultrasonics MX-5 Ultrasonic Thickness Gauge Dakota Ultrasonics Display Dakota Ultrasonics Transducer Making Measurements MX-5 Ultrasonic Thickness Gauge Condition and Preparation of Surfaces Performing a Probe-Zero Probe ZeroCalibration Calibration to a known thickness Calibration to a known velocity Two Point Calibration Moving series of dashes instead of a thickness value. When Scan ModeUsing the Beeper Alarm ModeDifferential Mode RS232 Serial Port Connecting To a Computer Transducer Selection Dakota Ultrasonics MX-5 Ultrasonic Thickness Gauge Blank Appendix a Product Specifications PhysicalBlank Application Notes ∙ Measuring pipe and tubing Appendix B∙ Measuring laminated materials MX-5 Ultrasonic Thickness Gauge Blank Sound Velocities of some Common Materials Appendix CBlank ∙ Warranty Statement ∙ Warranty InformationSONOTECH, INC