Delta DO 9704 manual Various operations

Page 23

ENGLISH

measuring unit.

*Press the Unit A or B key to select the temperature measuring unit in °C or °F.

4.Switching off the instrument.

*The instrument is switched off by pressing the ON/OFF key.

*Because of the Auto Power Off function the instrument may switch itself off during measure- ments. In this case press the ON/OFF key to switch it on again.

*The instrument usually switches off automatically after 8 minutes of inactivity, with the following exceptions:

a)Instrument in RCD status.

b)Instrument in Auto Power Off Disabled status.

In these two cases the instrument switches off only when the ON/OFF key is pressed.

c)Instrument during unloading of stored data.

d)Instrument during Serial Output of immediate data.

e)Instrument in storage function.

In the last three cases the instrument switches off automatically 8 minutes after the low battery warning and interrupts storage and unloading of data.

When it switches on again there are two possibilities:

1.If the battery is definitively low, even when the instrument switches on again LOU appears on the display together with the battery symbol.

The PROG key (P1+ENTER) enables the activation of the unloading of stored data even when the battery is low.

2.If the battery has had time to recover and when the instrument switches on its charge seems, even just a little, higher than the minimum value, LOU appears on the display without the bat- tery symbol for a short period of time (about 4 seconds), after which the instrument returns to normal operation; this is to remind the user that the instrument was previously in storage func- tion and that this procedure was interrupted by the low battery warning.

5.Various operations.

* For operations such as HOLD display, relative measurements, RCD storage, DATA CALL, Serial Output, storage in the internal memory, unloading of stored data, erasing of the internal memory, setting of the working parameters, see the description of the instrument keyboard.

Attention when using the keyboard

The use of the keys is relatively simple, but care must be taken to avoid setting it by mistake in an undesired mode. Ensure that HOLD, RCD, REL, MAX, MIN, MED, Serial Out/Memory are not displayed during normal operation.

87

Image 23
Contents Instructions Manual Page Pressure and Temperature DATA-LOGGER English Introduction Numbers Keyboard KEYREL Function DescriptionSimbols LIT Besides NumbersSerial Output Simbols LIT Besides Function Numbers DescriptionTom of the display 600 Baud P0+Press the Prog key to move on to step P1 300 BaudP1+ = data dump This parameter is used for the fol- lowing two functions P1+Storage Hours minutes SecondsFunction P4+ = set year P3+P5+ = set month P6+ = set day P7+ = set hours P8+ = set minutes Self cut-out P9+Complete printing Reduced printing P10+To stop storage, press the Enter key P10+ Start-Stop P11+P11 Data CallChes itself off again Operating modeWhen this operative KEY Simbols LIT BesidesSimbols LIT Besides Function Numbers Description Enter If the battery symbol does not flaAfter having taken note Restart itMore than 30,000 acquisitions Probe Connection HOW to MeasureVarious operations Method of Using the Instrument and Warnings Temperature MeasurementLOU Instrument Signals and MalfunctionsErr FULLOW Battery Warning and Battery Replacement Serial Interface RS 232C Storage of the InstrumentDumping data from the memory Command Type ReplyParity None Bit Stop bit Protocol Xon/Xoff Type of terminalMemory clear Printer Diagram doExample of a Report Printed by the Instrument English Software connection with Windows Hardware connectionOK to confirm the setting one click Mark local Echo one click 100 101 102 Switch on the measuring instrument103 OK to confirm Cancel in the next window104 Parity None105 106 107 108 109 110 To end receiving data from an instrumentSelect END from the pull-down sub-menu one click To quit running the Hyper TerminalVdc=465.5 Calibrating the Instrument doFlow Rate Measurements 113 Probe connection diagram Instrument doConnection of Probes to the Pressure Gauge Data Logger do Between 4 SEL R100 105 110 115 Turbine probe for measuring flow rateGauge Classification of Pressure MeasurementsConversion Factors and Tables = 2,491 x 102 Pa Psi = PaPsi = kPa = 3,386 x 103 Pa118 Minimum, Mean, Relative, A-B differential Instrument Technical DataConnector DIN 45326 8-pole Weight / dimensions 320 gr. / 215x73x38 mm120 121 Order Codes Guarantee 124 Examples of USE of the KeyboardSerial line Display blocked125 Pressing Data Call repeatedly givesInstrument programming 127 Sets yearSets day Sets hourData printing is reduced 128 Enables self cut-out +Normal operating Date and timeSerial output active Activation or disactivation of the serial outputActivation or disactivation of the Record function Function130 Activation and disactivation of the storage function131 Index132 EN61000-4-2, EN61010-1 level Safety