The transmission parameters with which the instrument is supplied are:

* Baud Rate

19200 Baud

* Parity

None

* N. bit

8

* Stop bit

1

* Protocol

Xon/Xoff

The data transmission speed may be changed by pressing the PROG key on the instrument to alter the set-up parameter P3. The possible Baud Rates are: 19200, 9600, 4800, 2400, 1200, 600, 300. The other transmission parameters are fixed.

All the messages reaching and leaving the instrument must be inserted in a “Communication fra- me” with the following structure:

Record-cr

 

 

 

where:

 

 

 

-Record-

constitutes the message

 

-cr-

Carriage Return (ASCII 0D)

 

HOST COMMANDS

 

 

COMMAND

TYPE

REPLY

AA

 

Type of terminal

MANOMETER – DL

AG

 

Firmware Version

Vx Rx

AH

 

Firmware Date

dd/mm/yy

A*

 

 

?

SA

 

Request

Input A

SB

 

Request

Input B

SC

 

Request

Input A-B

SI

 

Measurement

MANOMETER - DL

S*

 

 

NC

It is possible to unload data onto a computer using the Xon/Xoff protocol to control the data flow: The character Xoff (hex 13) on the serial line stops the data unloading in progress.

The character Xon (hex 11) reactivates the suspended data unloading in progress.

Dumping data from the memory

Press the PROG key twice until P1 appears on the display. Press the ENTER key.

The instrument continues to show P1 on the display and dumps the data from the instrument’s internal memory through the serial line.

92

Image 28
Contents Instructions Manual Page Pressure and Temperature DATA-LOGGER English Introduction Keyboard KEY NumbersFunction Description Simbols LIT BesidesNumbers RELTom of the display Simbols LIT Besides Function Numbers DescriptionSerial Output P0+ Press the Prog key to move on to step P1300 Baud 600 BaudP1+ = data dump P1+ StorageHours minutes Seconds This parameter is used for the fol- lowing two functionsFunction P3+ P4+ = set yearP5+ = set month P6+ = set day P7+ = set hours P8+ = set minutes P9+ Self cut-outP10+ To stop storage, press the Enter keyP10 Complete printing Reduced printingP11+ P11Data Call + Start-StopOperating mode When this operativeKEY Simbols LIT Besides Ches itself off againIf the battery symbol does not fla Simbols LIT Besides Function Numbers Description EnterMore than 30,000 acquisitions Restart itAfter having taken note HOW to Measure Probe ConnectionVarious operations Temperature Measurement Method of Using the Instrument and WarningsInstrument Signals and Malfunctions ErrFUL LOULOW Battery Warning and Battery Replacement Storage of the Instrument Serial Interface RS 232CCommand Type Reply Parity None Bit Stop bit Protocol Xon/XoffType of terminal Dumping data from the memoryMemory clear Diagram do PrinterExample of a Report Printed by the Instrument English Hardware connection Software connection with WindowsOK to confirm the setting one click Mark local Echo one click 100 101 Switch on the measuring instrument 102OK to confirm Cancel in the next window 103Parity None 104105 106 107 108 109 To end receiving data from an instrument Select END from the pull-down sub-menu one clickTo quit running the Hyper Terminal 110Calibrating the Instrument do Vdc=465.5Flow Rate Measurements Connection of Probes to the Pressure Gauge Data Logger do Probe connection diagram Instrument do113 SEL R 100 105 110 115Turbine probe for measuring flow rate Between 4Classification of Pressure Measurements GaugeConversion Factors and Tables Psi = Pa Psi = kPa= 3,386 x 103 Pa = 2,491 x 102 Pa118 Instrument Technical Data Connector DIN 45326 8-poleWeight / dimensions 320 gr. / 215x73x38 mm Minimum, Mean, Relative, A-B differential120 121 Order Codes Guarantee Examples of USE of the Keyboard Serial lineDisplay blocked 124Pressing Data Call repeatedly gives 125Instrument programming Sets year Sets daySets hour 127Enables self cut-out + Normal operatingDate and time Data printing is reduced 128Activation or disactivation of the serial output Activation or disactivation of the Record functionFunction Serial output activeActivation and disactivation of the storage function 130Index 131132 Safety EN61000-4-2, EN61010-1 level