Omega Engineering i.BTX-M manual Remote Access

Page 29

Port: (default 2000) is the default TCP port number for the port to which the sensor is connected. Ports 1000 (used for HTTPget, refer to Section 4.5), 2002, 2003, and 2004 are reserved for internal use.

Terminal Server usually describes a device that exchanges data between Ethernet/TCPIP networks and RS-232/RS-485 systems. With this iServer, the data is obtained digitally from the sensor (irrelevant to RS-232 or RS-485 interface) and can be accessed from anywhere on the network.

A computer program, such as NEWPORT’s Mail Notifier, OPC Server, iLOG, iFLASH or httpget can send TCP requests and obtain readings using the Terminal Server feature.

H)Remote Access

Remote IP Address: iServer can establish a connection to a remote device (e.g. an iLD Remote Display with an Ethernet iServer embedded board) with this IP.

Remote Port: (default 2000) the remote port number for the connection. Ports 1000 (used for HTTPget, refer to Section 4.5), 2002, 2003, and 2004 are reserved for internal use.

Remote Access:** Remote Access can be enabled and disabled. If enabled, the iServer can send its data to a remote node on the same network (the “Remote IP address” and “Remote Port” must be entered).

*TCP/UDP: when UDP mode is selected, Remote Access should be disabled and Remote IP and Port are the UDP remote listening IP and Port. If the Remote IP is set to 255.255.255.255, the UDP packet becomes a broadcasting packet which will allow any device listening to the Remote port to receive the packet.

**If Remote Access is enabled, Terminal Server is automatically disabled.

25

Image 29
Contents User’s Guide Servicing North America OMEGAnet On-Line Service Internet e-mail Info@omega.comTable of Contents IServer and iLD Big Display on the Ethernet Network List of FiguresFeatures Part Introduction Computer with Standard Web Browser Mounting DIP SwitchesPart 2 Hardware Parts of the iServer Unit Parts of iServer Unit Parts of the iServer UnitOpening the Unit Disassembly InstructionRelay Wiring Connections 2 10Base-T Crossover WiringRunning on Battery Power Network Communication Interfaces 1 10Base-T RJ-45 PinoutEthernet MAC Address Part Network ConfigurationNetwork Protocols IP Address DhcpExample C\Telnet 192.168.1.200 Default IP AddressPort Number Part Operations Testing the ConnectionICONNECT Software Assigning an IP Address using iCONNECTAccess Control Setting a New IP Address over the NetworkIServer Home Setup and Operation using the iServer WebRead Sensor Read Sensor Java Runtime Environment 1.4 Setup instructionsDiagnostics Java Runtime Environment 1.5 5.0 Setup instructionsAccessing iServer units using the internet Pressure/Temp Adjustable ChartOption Retrieving Data from FlashHost Name Refer to .4, DNS 10 Access Control#1 RTC is not set and Flash Card is empty Configuration#2 RTC is set, and recording is in progress #4 RTC is not set and the Flash Card is not emptySensor/Device Setup 11 ConfigurationTHi/Lo BHi/Lo BHi/Hi BLo/Hi BLo/Lo Real Time ClockFlash Recording Terminal Server Remote Access Remote End Char Sensor ParametersT75.7F P1014.9mbar Httpget Program Telnet SetupResponse Httpget using PortHttpget -r -S 00000000 Where Httpget and ARP to setup Device IP AddressApr -s 192.168.1.200 00-03-34-00-00-06-b6 ARP Protocol 14 ARP Commands and Responses15 iLOG Software Logging Data ILOG SoftwareMail Notifier Software InstallationEmail Connection Mapi Access 17 iServer Mail Notifier Profile Setup18 iServer Mail Notifier Device Setting Device Setting and Configuration Device setup requiresTemperature T Part 5 SpecificationsPower PackagingNetwork Interface Remote Access TunnelingAppendix a Glossary Appendix B IP Address Mask Dotted Notation Mask Binary Mask Bits Class a Appendix C IP NetmaskClass B Class CAscii Appendix D Ascii ChartDec Hex Binary Dec Hex Binary Char No Parity No parityAscii Control Codes Appendix D Ascii Chart ContinuationDec Hex Ctrl Key Definition Dec Hex Ctrl Key Definition Char EquivMeasurement Category Part Approvals Information CE ApprovalWARRANTY/DISCLAIMER Shop on line at omega.com