6.0PRESSURE EFFECTS

6.1Standard Transducers:

The Hastings Linear Mass Flowmeter can measure mass flow accurately without corrections for a variation in line pressure from one psia (.068 ATM) to 250 psig (18 ATMS.) Mass flow indications are possible with downstream pressures as low as .03 psia (.002 ATM), but the upstream must be much higher because of the increase in pressure drop across the transducer (see Section 8.0).

6.2Low Pressure Drop Transducers:

The low pressure drop Mass Flowmeter (using U-Series transducers) can measure flow accurately from .03 psia ( .002 ATM) to 15 psig ( 2 ATMS). The transducer is rated for a maximum pressure of 250 psig ( 18 ATMS). Mass flow indications are possible with downstream pressure as low as .1 Torr with upstream pressures of less than 30 Torr. This flowmeter has been successful for many years in measuring uranium hexafluoride flow rate under vacuum conditions, and has also been useful in measuring vacuum pumping speeds and in atmospheric sampling trains.

6.3High Pressure Transducers (Optional):

Hastings transducers having a P suffix have been pressure tested to 1500 psig using dry nitrogen gas. The transducer was cycled between 0 and 1500 psig for two periods with a minimum of 5 minutes each, and held at the pressure for a combined test period of 10 minutes. It was then leak tested on a Helium Leak Detector where a leak rate of no more than 3 X 10-6 SCCS was required.

The flowmeter has been designed and tested as described above for line pressure up to 1500 psig under normal usage. High pressure gas is always potentially dangerous and we strongly urge that extreme caution be taken in locating, installing, and operating this equipment.

Manual 111-082010_Nall Mass Flowmeters

Page 14 of 23

Image 14
Contents Teledyne Hastings Manual Print History Table of Contents 12.0 Features General InformationOperating Principle Operating PrincipleReceiving and Inspection Installation Instructions TransducerCables Power SupplyFilters Electrical Zero Check Using the Flowmeter Gas Mixtures Output SignalFlowmeter Factory Calibrated for a Special Gas Over-range AccuracyRepeatability Low Pressure Drop Transducers Pressure EffectsStandard Transducers High Pressure Transducers OptionalPower Supply Temperature EffectsAmbient Temperatures Transducer Gas TemperatureChanges in DP with Changes in Line Pressure Differential PressureTypical Pressure Drops Increase in Line PressureDecrease in Line Pressure DP Increase Due to Fouling of the TransducerCalibration Procedures CalibrationTransducer Power Supply Troubleshooting MaintenanceTransducers Cleaning Damage to the TransducerDouble Point Relay Flowmeter/Alarm DNALL-P AccessoriesOptions and Accessories Totalizers TNALL-P DescriptionsCurrent Converter NAll-P/CC MaintenanceChassis Model NALL-C Non-Warranty Repair Policy Warranty and RepairWarranty Repair Policy