Teledyne instruction manual Maintenance, Chassis Model NALL-C, Current Converter NAll-P/CC

Page 22

a)What is the sensitivity of the totalizer?

b)What is the total flow indicated by the 13,500 count?

c)What is the average flow rate over the 7 hour period?

Solution: a) The full scale flow rate for the TNALL-P/HS-10S is 10 SCCM. The full-scale count rate for the totalizer is 100 counts per minute.

Sensitivity

= F. S. flow rate

 

F. S. countrate

=10 std. cm³/min.

100 counts/min.

d)The total flow indicated by 13,500 counts is:

Total flow = total counts X sensitivity

=13,500 counts X 0.1 std cm³/min.

=1350 std cm³

c)The average flow rate over the 7 hour period is the total flow divided by the time:

Average Flow Rate = Total Flow

Time

=1350 std cm³

7 hrs X 60 min/hr

=1350 std cm³

420 min.

=3.2 std cm³/min.

If a special gas is used with the flowmeter (other than that for which it was originally calibrated), refer to Section 5.5 to determine the new full-scale flow rate for the special gas. Once the full-scale rate for the special gas is obtained, it may be used with the equations above to determine the sensitivity of the totalizer in std. cm³/count.

11.2.4 Maintenance

The totalizer is all solid-state and is designed for a long life expectancy and no maintenance is normally required. Should the totalizer or counter need repair, return them to Teledyne Hastings with a statement describing the problem; use the form on page 35.

11.3Chassis Model (NALL-C):

Hastings Chassis Linear Mass Flowmeters are designed for installations where the electronic circuits are remotely mounted from both the transducer and readout. It is particularly useful in applications where the meter must be located long distances from the transducer. Thus, the transducer cable is kept short, and the meter may be located long distances from the Chassis without concern for cable losses. Also, panel space is reduced to that of the meter only. The Chassis is designated by adding the suffix “C” to the Cabinet Model number; NALL becomes NALL-C.

11.4Current Converter (NAll-P/CC):

A 4-20 mA current converter is an option available with the Hastings Mass Flowmeter. The Hastings 4-20 mA may be built into a flowmeter package or supplied as a separate unit. The converter produces a 4-20 mA current signal from the 0-5 VDC output of the flowmeter.

Manual 111-082010_Nall Mass Flowmeters

Page 22 of 23

Image 22
Contents Teledyne Hastings Manual Print History Table of Contents 12.0 Features General InformationOperating Principle Operating PrincipleReceiving and Inspection Installation Instructions TransducerFilters Power SupplyCables Electrical Zero Check Using the Flowmeter Flowmeter Factory Calibrated for a Special Gas Output SignalGas Mixtures Repeatability AccuracyOver-range Low Pressure Drop Transducers Pressure EffectsStandard Transducers High Pressure Transducers OptionalPower Supply Temperature EffectsAmbient Temperatures Transducer Gas TemperatureChanges in DP with Changes in Line Pressure Differential PressureTypical Pressure Drops Increase in Line PressureDecrease in Line Pressure DP Increase Due to Fouling of the TransducerCalibration Procedures CalibrationTransducer Power Supply Troubleshooting MaintenanceTransducers Cleaning Damage to the TransducerDouble Point Relay Flowmeter/Alarm DNALL-P AccessoriesOptions and Accessories Totalizers TNALL-P DescriptionsChassis Model NALL-C MaintenanceCurrent Converter NAll-P/CC Warranty Repair Policy Warranty and RepairNon-Warranty Repair Policy