10.0Maintenance
10.1Transducers:
10.1.1 Cleaning
With proper care in installing and use, the transducer will require little or no maintenance. Should the small passages in the transducer become partially clogged, they can be cleaned with any suitable solvent such as acetone, vythene, etc. and/or blown out with clean air under moderate pressure (up to 50 psig).
NOTE: If solvents are used for cleaning the transducer, sporadic indications may occur when it is returned to service due to residual moisture in the passages of the transducer. Time should be allowed for the liquid solvent to evaporate.
10.1.2 Damage to the Transducer
The transducer can be checked for internal electrical damage by disconnecting the cable and using an ohm- meter to measure resistances at the
10.2Power Supply:
NALL and
10.2.1 Power Supply Troubleshooting
If a problem does occur, the transducer heating voltage and the transducer output can be checked (refer to Figure 9.1). To check the heating voltage, the flowmeter must be connected to the transducer and a true R.M.S. voltmeter must be used. With the TRMS voltmeter connected between test points D and A (TRMS AC test), the voltage should be 16.00 TRMS VAC +/- 0.1 VAC (and can be adjusted by the BIAS pot, R6). The transducer output is measured between test points BC and E (millivolt test) using a DC millivolt meter. This output will vary as flow through the transducer varies. At full scale flow for the transducer, the millivolt meter should read between 0.75 and 7 millivolts DC, depending on the individual transducer. This signal is amplified to produce the flowmeter output.
If it is necessary to return the flowmeter for repair, please return the transducer, power supply, and cable to Teledyne
Manual | Page 20 of 23 |