Example: | (A) | What is the output sensitivity of | ||
| (B) What is the flow rate for a 2.71 volt output? | |||
Solution: | (A) | Sensitivity | = | rated flow |
|
|
|
| Rated output |
|
|
| = | 1000 SCCM |
|
|
|
| 5.00 VDC |
|
|
| = | 200 SCCM/VDC |
| (B) | Mass Flow | = | output X sensitivity |
|
|
| = | 2.71v X 200 SCCM/VDC |
|
|
| = | 542 SCCM |
5.7Accuracy:
The accuracy of the flow indication is +/- 1% of full scale, with the exception of the
5.8Repeatability:
The repeatability over a
clean, dry system. Under reasonable constant conditions, a repeatability of +/- 1 to+/- 2% of INDICATION can be expected on a
5.9Over-range:
The flowmeter has an output signal which is directly proportional to mass flow and linear to +/- of full scale from zero flow to the normal
The flowmeter can be used to measure flow rates higher than the rated maximum flow, but the output signal normally becomes