4.0INSTALLATION INSTRUCTIONS.

4.1TRANSDUCER

4.1.1Orientation of the Transducer

The transducer may be mounted in any position, as long as the direction of gas flow through the transducer is from “IN” to “OUT” as marked on the transducer base. (U-Series transducers should be mounted in a horizontal position only.)

4.1.2Mounting the Transducer

There are two ¼-20 threaded holes 3/8” deep in the bottom of the transducer that can be used to secure it to a mounting bracket, if desired. When the transducer is used in combination with an L or LU Series Laminar Flow Element (LFE), the LFE should be supported instead of the transducer to prevent undue strain on the connectors between the transducer and the LFE. Standard pipe support rings or pipe hangers are usually satisfactory for supporting the LFE.

4.1.3Inlet and Outlet Connections

The table 4.1 describes the inlet and outlet connections for all standard transducers. (If it is necessary to reduce the pipe size or install an elbow on either side of an L or LU Series Laminar Flow Element, it is recommended that a straight pipe 12” in length, and of pipe or flange size stated below, be connected directly to the LFE before connecting a smaller diameter pipe or an elbow.)

4.1.4Sealing the Threaded Connections

Many users find that Teflon tape is an excellent sealant for most applications, however, any sealant material compatible with the flow system is acceptable. Caution must be exercised during assembly and disassembly of the threaded connections to prevent shreds of the Teflon tape, or the sealant from entering the flow line, where they could block the small passages in the transducer.

Manual 111-082010_Nall Mass Flowmeters

Page 8 of 23

Image 8
Contents Teledyne Hastings Manual Print History Table of Contents 12.0 Features General InformationOperating Principle Operating PrincipleReceiving and Inspection Installation Instructions TransducerCables Power SupplyFilters Electrical Zero Check Using the Flowmeter Gas Mixtures Output SignalFlowmeter Factory Calibrated for a Special Gas Over-range AccuracyRepeatability Pressure Effects Standard TransducersLow Pressure Drop Transducers High Pressure Transducers OptionalTemperature Effects Ambient Temperatures TransducerPower Supply Gas TemperatureDifferential Pressure Typical Pressure DropsChanges in DP with Changes in Line Pressure Increase in Line PressureDecrease in Line Pressure DP Increase Due to Fouling of the TransducerCalibration Procedures CalibrationTransducer Maintenance Transducers CleaningPower Supply Troubleshooting Damage to the TransducerAccessories Options and AccessoriesDouble Point Relay Flowmeter/Alarm DNALL-P Totalizers TNALL-P DescriptionsCurrent Converter NAll-P/CC MaintenanceChassis Model NALL-C Non-Warranty Repair Policy Warranty and RepairWarranty Repair Policy