Teledyne NALL instruction manual Calibration Procedures

Page 18

9.0Calibration

9.1CALIBRATION PROCEDURES

The Hastings Mass Flowmeter has been carefully inspected and calibrated at the factory before shipment and will give long, reliable service. Calibration is stable, and recalibration is seldom necessary under normal operating conditions; however, to maintain optimum accuracy it is advisable to check calibration on an annual basis. The unit may be returned to the factory for this purpose, but if that is undesirable, the following calibration procedure may be of assistance:

1.Connect the power supply to the proper power source (Section 3.4).

2.Turn the power supply “ON” and allow 30 minutes for warm-up.

3.Set the “electrical zero” (Section 3.5).

4.Connect the inlet side of the transducer through a metering valve, to a well-regulated air source and the outlet to a reliable flow reference such as the Hastings Mini-Flo Calibrator.

5.Check transducer heating voltage as outlined in Section 10.2.1.

6.During normal calibration, a flowmeter is spanned at 80% of full scale to obtain best average linearity.

Example: A 50 SCCM flowmeter with 0-5 VDC output at the binding posts is to be calibrated. If a reference flow standard is set at a 40 SCCM flow rate, the flowmeter should indicate 80% (.80 X 50 SCCM

=40 SCCM) of full scale. At this flow rate, the voltage at the binding posts should read 4 VDC (80% X 5 VDC).

CAUTION: Most flow standards are volumetric devices and must be corrected for temperature and pressure to standard conditions of 0oC and 760 Torr. This correction amounts to several percent at normal room temperatures and pressures and should not be neglected.

7.If the flowmeter does not agree with the reference flow standard, it is first necessary to correct the output at the binding posts. This is done by making a small adjustment to the “GAIN” potentiometer R-30, located on the circuit board inside the power supply. This adjustment will also produce a change in the meter indication.

8.If the meter reading (flow) does not correspond with the binding post’s output (voltage), make a small adjustment to the “METER” trim potentiometer R-32, located on the circuit board inside the power supply. This correction has no effect on the output voltage.

9.PC-581A Board Only: Refer to Figure 9.1. Do not change the dipswitch (S2) settings. These settings are factory pre-set and should only need changing if there is a change in the transducer or the units of calibration. Switches 1, 2, & 3 change gain and 5, 6, 7, & 8 change the meter range to engineering units. The number 4 position should remain closed.

Manual 111-082010_Nall Mass Flowmeters

Page 18 of 23

Image 18
Contents Teledyne Hastings Manual Print History Table of Contents 12.0 Features General InformationOperating Principle Operating PrincipleReceiving and Inspection Installation Instructions TransducerPower Supply FiltersCables Electrical Zero Check Using the Flowmeter Output Signal Flowmeter Factory Calibrated for a Special GasGas Mixtures Accuracy RepeatabilityOver-range Low Pressure Drop Transducers Pressure EffectsStandard Transducers High Pressure Transducers OptionalPower Supply Temperature EffectsAmbient Temperatures Transducer Gas TemperatureChanges in DP with Changes in Line Pressure Differential PressureTypical Pressure Drops Increase in Line PressureDecrease in Line Pressure DP Increase Due to Fouling of the TransducerCalibration Procedures CalibrationTransducer Power Supply Troubleshooting MaintenanceTransducers Cleaning Damage to the TransducerDouble Point Relay Flowmeter/Alarm DNALL-P AccessoriesOptions and Accessories Totalizers TNALL-P DescriptionsMaintenance Chassis Model NALL-CCurrent Converter NAll-P/CC Warranty and Repair Warranty Repair PolicyNon-Warranty Repair Policy