11 APPENDIX B, EMC INFORMATION
GUIDANCE AND MANUFACTURER’S DECLARATION
Electromagnetic Emissions—for all Equipment and Systems
Guidance and manufacturer’s declaration— electromagnetic emission
The Grafco Ultrasound Pocket Doppler is intended for use in the electromagnetic environment specified below. Do not use the system in environments which do not comply with the specifications listed below.
Emission | Compliance | Electromagnetic environment– |
test |
| guidance |
RF emission | Group 1 | The Grafco Ultrasound Pocket Doppler |
CISPR 11 |
| uses RF energy only for its internal function. |
|
| Therefore, its RF emissions are very low |
|
| and are not likely to cause any interference |
|
| in nearby electronic equipment. |
RF emission | Class B | The Grafco Ultrasound Pocket Doppler |
CISPR 11 |
| is suitable for use in all establishments, |
|
| including domestic establishments and |
|
| those directly connected to the public |
|
| |
|
| supplies buildings used for domestic |
|
| purposes. |
Electromagnetic
Guidance and manufacture’s
The Grafco Ultrasound Pocket Doppler is intended for use in the electromagnetic environment specified below. Do not use the system in environments which do not comply with the specifications listed below.
Immunity | IEC 60601 | Compliance | Electromagnetic |
test | test level | level | |
Electrostatic | ±6 kV | ±6 kV | Floors should be wood, |
discharge | contact ±8 | contact ±8 | concrete or ceramic tile. If |
(ESD) IEC | kV air | kV air | floor is covered with synthetic |
|
| material, the relative humidity | |
|
|
| should be at least 30% RH. |
| 36 |