WARNING: Do not the use Pocket Doppler if the bat- tery cover is open.

WARNING: Use only those probes provided by GF Health Products, Inc. intended to be used with the Grafco Pocket Doppler.

WARNING: Do not stretch the probe cable further than 6.5 feet (2 meters).

WARNING: Electromagnetic Interference: Before op- eration, ensure that the environment in which the de- vice is operated is not subject to any source of strong electromagnetic interference, such as radio transmit- ters, mobile telephones, etc.

WARNING: The following safety checks/tests must be performed at least once every two years or as speci- fied in your facility's test and inspection protocol, by a qualified person with adequate training, knowledge, and practical experience:

Inspect the equipment for mechanical and func- tional damage.

Ensure the safety labels are legible.

Verify that the device functions properly as de- scribed in this manual.

Test according to the pregnant woman's leakage current (IEC 60601-1/1988: Limit: 100 uA (B)).

The leakage current should never exceed the speci- fied limit. Record the data and store with the Grafco Pocket Doppler. If the device is not functioning prop- erly or fails any of the above tests, contact GF Health Products, Inc. Technical Support at the phone number on the back cover, or your GF authorized distributor.

14423-INS-LAB-RevD11

7

Image 7
Contents Ultrasound Pocket Doppler Contents Introduction Included FeaturesIntended USE of this Device Product Description Optional AccessoriesSafety Precautions Safety InformationSafety Guidance INS-LAB-RevD11 INS-LAB-RevD11 INS-LAB-RevD11 Symbols GF Health PRODUCTS, INC. Freight Policy HandlingUnpacking Inspection StorageAppearance and Features Ultrasound Pocket Doppler and AccessoriesButton PowerSocket Jack Definition Attachment Battery Rear Panel Feature Function Pocket ClipSocket Name Picture Top Panel FeatureMode Button Left Panel Feature Function Volume ControlPocket Doppler WATER-RESISTANT Probes BatteryInstalling or Replacing the Battery SetupINS-LAB-RevD11 Placing the Probe in the Pocket Doppler OperationProbe Operation Removing the Probe from the Pocket DopplerSwapping Probes Turning on the Pocket DopplerFH Monitoring with 2.0 MHz Water-Resistant Probe Fetal Heart FH MonitoringVascular Monitoring with 4.0 or 8.0 MHZ Probe Optional Turning OFF the Pocket Doppler Replacing the BatteryMaintenance MaintenanceCleaning the Pocket Doppler CleaningCleaning the Pocket Doppler Probe Disinfecting the Pocket Doppler Probe DisinfectionSafety SpecificationsPhysical Power BatteryUltrasound Limited Warranty INS-LAB-RevD11 GF Health Products, Inc. Graham-Field EC Declaration of Conformity Appendix A, EC Declaration of ConformityGuidance and MANUFACTURER’S Declaration Appendix B, EMC InformationRF IEC DistanceRated Separation distance according to Recommended Separation DistancesFrequency of transmitter m INS-LAB-RevD11 Overall Sensitivity 2 MHz Probe Appendix C, Overall Sensitivity=Ad+B+C Index INS-LAB-RevD11