The span gas cylinder must be equipped with a pressure regulator with a metallic diaphragm. The regulator should be set to approximately
Allow the span gas to pass through the O2 sensor until the reading has stabilized (approx. 10 min.). The O2 transmitter can then be adjusted (see instructions elsewhere in this section) to match the analyzed content of the span gas. When the calibration is complete, return the selector valve (a) to the sample position and readjust the flow valve (c), (if necessary) so that flow meter
(b) reads 0.4 scfh.
Figure 3-5 Sample System Component Locations