Teledyne operating instructions Introduction Model OT-2 System

Page 6

1 Introduction

Model OT-2 System

 

 

 

 

 

 

•Accurate (±2% of full scale at constant temperature on 100 ppm, or 1000 ppm scale; ±1 ppm on 0–10 ppm scale at constant temperature)

•Insensitive to flow variations

•Fast response and recovery

•Long life, maintenance-free Micro-Fuel Cell oxygen sensor

•Unaffected by reducing agents (HC's, CO, SO2, etc.)

•Easy to calibrate, no zero gas required

•12VDC Battery Powered w/ 1-5 VDC Output

•Rugged NEMA 4, bulkhead mounted enclosure

•Stainless Steel sample system (nylon cell holder)

•H2S scrubber

•Pressure regulation with gage

•Sample filter

•Flowmeter

•Heated cell holder assembly

1.3Typical Applications

Although the OT-2 System is designed in conjunction with the Elsag Baily Total Flow TM System for natural gas production, it also is capable of monitoring oxygen at the ppm level in a variety of gases. A few typical applications include:

•Gas gathering and transportation

•Monitoring inert gas blanketing

•Air separation and liquefaction

•Chemical feedstock analysis

•Petrochemical process control

•Heat treating and bright annealing processes

•Quality assurance

•Gas certification

•Welding applications

1-2

Teledyne Analytical Instruments

Image 6
Contents Model OT-2 Oxygen Transmitter System Model OT-2 System Copyright 1999 Teledyne Analytical InstrumentsTable of Contents Ground Model OT-2 System Introduction Introduction Model OT-2 System Front View Introduction Model OT-2 System Model OT-2 System Operational Theory IntroductionOperational Theory Model OT-2 System Micro-Fuel Cell Sensor Principles of OperationCross Section of a Micro-Fuel Cell not to scale Electrochemical Reactions Model OT-2 System Operational Theory Characteristic Input/Output Curve for a Micro-Fuel Cell Electronics and Signal Processing Sample Flow Description See Figure Piping Layout Operational Theory Model OT-2 System Installation Model OT-2 System OperationOperation Model OT-2 System Mounting DetailsGas Connections Operation Model OT-2 System Front View of Transmitter Operation Model OT-2 System Installing the Micro-Fuel Cell CalibrationSample System Component Locations Span Gas % F.S. x 4 + = Output in VDC 100 Testing the System 3 H2S ScrubberOperation Model OT-2 System Model OT-2 System Maintenance Model OT-2 System Scrubber Replacement Maintenance Model OT-2 System Appendix Qty Part Number Description Section I Product Identification Section III Physical Hazards Section V Emergency and First Aid Procedures Appendix