Agilent Technologies 54600-Series manual Trigger System, Pulse Width

Page 16

Performance Characteristics (continued)

Horizontal (continued)

Digital Delta-t Accuracy

54621A/D, 54622A/D, 54624A:

(non-Vernier settings)

Same Channel: ±0.01% reading ±0.1% screen width ±(1 digital sample period, 2.5 or 5 ns

 

based on sample rate of 200/400 MSa/s)

 

Channel-to-Channel: ±0.01% reading ±0.1% screen width ±(1 digital sample period, 2.5 or 5 ns)

 

±chan-to-chan skew (2 ns typical, 3 ns maximum)

 

54641A/D, 54642A/D:

 

Same Channel: ±0.005% reading ±0.1% screen width ±(1 digital sample period, 1ns)

 

Channel-to-Channel: ±0.005% reading ±0.1% screen width ±(1 digital sample period) ±chan-to-chan skew

 

Same Channel Example (54641A/D, 54642A/D):

 

for signal with pulse width of 10 µs, scope set to 5 µs/div (50 µs screen width),

 

delta-t accuracy = ±{.005%(10 µs) + 0.1%(50 µs) + 1 ns} = 51.5 ns

 

 

Delay Jitter

<1 ppm

 

 

RMS Jitter

0.025% screen width + 30 ps

 

 

Modes

Main, Delayed, Roll, XY

 

 

XY

Bandwidth: Max bandwidth

 

Phase error @ 1 MHz: 1.8 degrees

 

Z Blanking: 1.4 V blanks trace (use external trigger) - 54621A/D, 54622A/D, 54624A only

 

 

Trigger System

 

 

 

Sources:

54621A/622A, 54641A/642A: Ch 1, 2, line, ext

 

54621D/622D, 54641D/642D: Ch 1, 2, line, ext, D15 - D0

 

54624A: Ch 1, 2, 3, 4, line, ext

 

 

Modes

Auto, Triggered (normal), SingleAuto level (54621A/D, 54622A/D, 54624A only)

 

 

Holdoff Time

~60 ns to 10 seconds

 

 

Selections

Edge, Pulse Width, Pattern, TV, Duration, Sequence, CAN, LIN, USB, I2C, SPI

Edge

Trigger on a rising or falling edge of any source

 

 

Pattern

Trigger on a pattern of high, low, and don’t care levels and/or a rising or falling edge established across any of the sources.

 

The analog channel’s high or low level is defined by that channel’s trigger level.

 

 

Pulse Width

Trigger when a positive- or negative-going pulse is less than, greater than, or within a specified range on any of

 

the source channels.

 

Minimum pulse width setting: 5ns (2 ns on 54641A/D, 54642A/D analog channels)

 

Maximum pulse width setting: 10 s

 

 

TV

Trigger on any analog channel for NTSC, PAL, PAL-M, or SECAM broadcast standards on either positive or negative

 

composite video signals. Modes supported include Field 1, Field 2, or both, all lines, or any line within a field. Also supports

 

triggering on non-interlaced fields. TV trigger sensitivity: 0.5 division of synch signal.

 

 

Sequence

Arm on event A, trigger on event B, with option to reset on event C or time delay.

 

 

CAN

Trigger on CAN (Controller Area Network) version 2.0A and 2.0B signals. It can trigger on the Start of Frame bit of a data

 

frame, a remote transfer request frame, or an overload frame.

 

 

LIN

Trigger on LIN (Local Interconnect Networking) sync break at beginning of message frame.

 

 

USB

Trigger on USB (Universal Serial Bus) Start of Packet, End of Packet, Reset Complete, Enter Suspend, or Exit Suspend on

 

the differential USB data lines. USB low speed and full speed are supported.

 

 

* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

16

Image 16
Contents 60 to 500 MHz, up to 2 GSa/s Ideal for mixed analog and digital analysisMB to 8 MB MegaZoom deep memory standard Selection Guide Selection GuideModel Maximum Sample Rate Maximum Memory ChannelsChannel scope model 54624A Affordable and Responsive MegaZoom deep memoryStandard features Connectivity made easierBuilt-in floppy drive Built-in help at your fingertipsOther standard features Deep memory transfer over the busN2758A can Trigger Module Flexible Triggering TriggeringCAN, LIN, and USB Triggering I2C and SPI Triggering Microcontroller DSP ApplicationsCapture all of your data at once Standard Serial Triggering Probes Agilent Passive Probes Selection GuideAcquisition Analog Channels Performance CharacteristicsVertical System Analog Channels 54641A/D, 54642A/DDC to max bandwidth 40 dB Horizontal Same Channel Example 54641A/D, 54642A/DTrigger System Pulse WidthI2C External EXT TriggeringMeasurement Features Display SystemGeneral Characteristics StoragePower Requirements OtherOrdering Information Warranty and Calibration Options Accessories DescriptionAvailable Options Description Passive Probes DescriptionActive probes Description Logic Probes DescriptionCables Description 5968-8152EN