Performance Characteristics (continued)

Display System

Display

7-inch raster monochrome CRT

 

 

Throughput of Analog Channels

25 million vectors/sec per channel with 32 levels of intensity

 

 

Resolution

255 vertical by 1000 horizontal points (waveform area) 32 levels of gray scale

 

 

Controls

Waveform intensity on front panel. Vectors on/off; infinite persistence on/off 8 x 10 grid with continuous intensity control

 

 

Built-in Help System

Key-specific help in 11 languages displayed by pressing and holding key or softkey of interest

 

 

Real Time Clock

Time and date (user setable)

 

 

Measurement Features

 

 

 

Automatic Measurements

Measurements are continuously updated. Cursors track current measurement

 

 

Voltage (analog channels only)

Peak-to-Peak, Maximum, Minimum, Average, Amplitude, Top, Base, Overshoot, Preshoot, RMS (DC)

 

 

Time

Frequency, Period, + Width, - Width and Duty Cycle on any channels

 

Rise time, Fall time, X at Max (Time at max volts), X at Min (Time at min volts), Delay, and Phase on analog channels only.

 

 

Counter

Built-in 5 digit frequency counter on any channel. Counts up to 125 MHz

 

 

Threshold Definition

Variable by percent and absolute value; 10%, 50%, 90% default for time measurements

 

 

Cursors

Manually or automatically placed readout of Horizontal (X, ∆X, 1/∆X) and Vertical (Y, ∆Y)

 

Additionally digital or analog channels can be displayed as binary or hex values

 

 

Waveform Math

One function of 1-2, 1*2, FFT, differentiate, integrate.

 

Source of FFT, differentiate, integrate: analog channels 1 or 2 (or 3 or 4 for 54624A), 1-2, 1+2, 1*2

 

 

FFT

 

 

 

Points

Fixed at 2048 points

 

 

Source of FFT

Analog channels 1 or 2 (or 3 or 4 on 54624A only), 1+2, 1-2, 1*2

 

 

Window

Rectangular, Flattop, Hanning

 

 

Noise Floor

-70 to -100 dB depending on averaging

 

 

Amplitude Display

In dBV, dBm@50 Ω

 

 

Frequency Resolution

0.097656/(time per div.)

 

 

Maximum Frequency

102.4/(time per div.)

 

 

* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

18

Image 18
Contents Ideal for mixed analog and digital analysis 60 to 500 MHz, up to 2 GSa/sMB to 8 MB MegaZoom deep memory standard Model Selection GuideSelection Guide Maximum Sample Rate Maximum Memory ChannelsChannel scope model 54624A Affordable and Responsive MegaZoom deep memoryBuilt-in floppy drive Standard featuresConnectivity made easier Built-in help at your fingertipsOther standard features Deep memory transfer over the busN2758A can Trigger Module Triggering Flexible TriggeringCAN, LIN, and USB Triggering I2C and SPI Triggering Microcontroller DSP ApplicationsCapture all of your data at once Standard Serial Triggering Probes Agilent Passive Probes Selection GuidePerformance Characteristics Acquisition Analog ChannelsVertical System Analog Channels 54641A/D, 54642A/DDC to max bandwidth 40 dB Horizontal Same Channel Example 54641A/D, 54642A/DTrigger System Pulse WidthI2C External EXT TriggeringMeasurement Features Display SystemGeneral Characteristics StoragePower Requirements OtherOrdering Information Available Options Description Warranty and Calibration OptionsAccessories Description Passive Probes DescriptionLogic Probes Description Active probes DescriptionCables Description 5968-8152EN